欢迎访问ic37.com |
会员登录 免费注册
发布采购

NLP110-9994 参数 Datasheet PDF下载

NLP110-9994图片预览
型号: NLP110-9994
PDF下载: 下载PDF文件 查看货源
内容描述: 中低功率AC / DC电源 [LOW TO MEDIUM POWER AC/DC POWER SUPPLIES]
分类和应用:
文件页数/大小: 3 页 / 171 K
品牌: ARTESYN [ ARTESYN TECHNOLOGIES ]
 浏览型号NLP110-9994的Datasheet PDF文件第2页浏览型号NLP110-9994的Datasheet PDF文件第3页  
NLP110 Medical Series
Single and triple output
LOW TO MEDIUM POWER AC/DC POWER SUPPLIES
80-110W AC/DC Universal Input Switch Mode Power Supplies
1
Provides low voltage outputs
Compliance to EN61000-3-2 (Power Factor = 0.98)
Smallest industry standard package 6.5 x 3.0 x 1.26 inches
90VAC to 264VAC universal input range
UL, cUL, and VDE safety approvals and CE-Marked to MDD
Overvoltage and short circuit protection
VDE0871-A, EN55022-A, EN55011-A conducted noise
VDE0871-A, EN55022-A, EN55011-A radiated noise
Meets all applicable and relevant immunity standards
EN61000-4-2,4,5 and 6
The NLP110 Medical is a series of 110W AC/DC open-frame power supplies in a
3” x 6.5” x 1.26” package. The NLP110 Medical is power factor corrected to meet EN61000-3-2
and is available in single and multiple output models. The multiple output models have a capability
that is unique to the power supply market - they can deliver equal amounts of current on multiple
low voltage outputs to support designs that will span the industry transition from 5V logic to 3.3V
logic. The NLP110 Medical is practically the same size as a 65 Watt 3” x 5” product, e.g. NLP65
Medical - the one difference is that the footprint is a mere 1.5 inches longer. The NLP110 Medical
series, with full medical safety approval per UL2601-1, C22.2 No. 601.1-M90, VDE0750,
IEC60601-1 and EN60601-1 and CE marked, greatly accelerating design-in time and reducing
system compliance costs.
All specifications are typical at nominal input, full load at 25°C unless otherwise stated
2 YEAR WARRANTY
SPECIFICATIONS
OUTPUT SPECIFICATIONS
Total regulation
(Line and load)
Rise time
Transient response
Main output
Auxiliary outputs
At turn-on
Main output
75% to 100%
step at 0.1A/µs
±2.0%
±5.0%
1.0s, max.
5.0% or 250mV
max. dev., 1ms max.
recovery to 1%
±0.02%/°C
Main outputs
Cyclic operation
Single and multiple
125%, ±10%
Continuous
See table
EMC CHARACTERISTICS (continued)
Surge
Fast transients
Radiated immunity
Conducted immunity
EN61000-4-5
EN61000-4-4
EN61000-4-3
EN61000-4-6
Level
Level
Level
Level
3
3
3
3
GENERAL SPECIFICATIONS
Hold-up time
Efficiency
Isolation voltage
Approvals and
standards
Weight
120VAC, 60Hz
120VAC, 80W (-9605)
Input/output
Input/chassis
16.7ms @ 110W
75% min.
4000VAC
1500VAC
Temperature coefficient
Overvoltage protection
Short circuit protection
Minimum output current
INPUT SPECIFICATIONS
Input voltage range
(See Note 6)
Input frequency range
Input surge current
(cold start)
Safety ground
leakage current
Input current
120VAC
230VAC
230VAC, 50Hz
120VAC
120VAC
230VAC
230VAC
@
@
@
@
80W
110W
80W
110W
Universal input
EN60601-1, IEC60601-1,VDE0750,
UL2601-1, C22.2 No. 601.1-M90
283g (10oz)
MIL-HDBK-217F
(9)
90 to 264VAC
47Hz to 63Hz
18A max.
35A max.
<100µA
0.9A
1.3A
0.48A
0.7A
rms
rms
rms
rms
MTBF (@ 25°C)
220,000 hours min.
ENVIRONMENTAL SPECIFICATIONS
Thermal performance
Operating ambient,
0°C to +50°C
(See derating curve)
Non-operating
-40°C to +85°C
50°C to 70°C, ambient,
Derate to
convection cooled
50% load
0°C to 50°C, ambient,
80W
convection cooled
0°C to 50°C ambient,
110W
150LFM forced air
Peak (0°C to +50°C, 60s) (See Note 3)
Non-condensing
Operating
Non-operating
5Hz to 500Hz
per MIL-STD-810E
5% to 95% RH
10,000 feet max.
30,000 feet max.
2.4G rms peak
516.4 Part IV
Input fuse
UL/IEC127
(12)
F3.15A H, 250VAC
Relative humidity
Altitude
EMC CHARACTERISTICS
Conducted emissions
Radiated emissions
Harmonic current
emission correction
ESD air
ESD contact
EN55022, FCC part 15
EN55022, FCC part 15
EN61000-3-2
EN61000-4-2
EN61000-4-2
Level A
Level A
Compliant
Level 3
Level 3
Vibration (See Note 8)
Shock
File Name: NLP110M.PDF Rev: 13 Dec 2002