欢迎访问ic37.com |
会员登录 免费注册
发布采购

LPS65-M 参数 Datasheet PDF下载

LPS65-M图片预览
型号: LPS65-M
PDF下载: 下载PDF文件 查看货源
内容描述: 60瓦,医疗 [60 Watts, Medical]
分类和应用: 医疗
文件页数/大小: 2 页 / 611 K
品牌: ASTEC [ Astec America, Inc ]
 浏览型号LPS65-M的Datasheet PDF文件第2页  
60 Watts
Medical
LPS60 Series
Total Power:
60-80 Watts
Input Voltage:
85-264 VAC
120-300 VDC
# of Outputs:
Single
Electrical Specs
Input
Input range
Frequency
Inrush current
Input current
Efficiency
EMI filter
85-264 VAC
120-300 VDC
47-440 Hz
<18 A peak @ 115 VAC;
<36 A peak @ 230 VAC,
cold start @ 25°C
1.5 A max. (RMS) @ 115 VAC
70% typical at full load
FCC Class A conducted
CISPR 22 Class A conducted
EN55022 class A conducted
VDE 0878 PT3 class A conducted
<75 µA @ 50/60 Hz, 264 VAC input
Special Features
Medical safety approvals
Universal input
3” x 5” footprint
Remote sense
Built-in EMI filter
Low output ripple
Adjustable output
Overvoltage protection
Overload protection
110 KHz switching frequency
LPX60 enclosure kit available
Safety ground leakage
current
Output
Maximum power
Adjustment range
Hold-up time
Overload protection
Overvoltage protection
Remote sense
60 W for convection;
80 W with 30 CFM forced air
-5, +10% minimum
20 ms @ 60 W load, 115 VAC nominal line
Short circuit protection on all outputs.
Case overload
protected @ 110-145%
above peak rating
5 V output: 5.7-6.7 VDC. Other outputs
10% to 25% above nominal output
Compensates for 0.5 V lead drop min. Will
operate without remote sense connected.
Reverse connection protected.
Environmental
Operating temperature: 0° to 50°C ambient
derate each output at 2.5% per degree from 50° to
70°C
Electromagnetic susceptibility: designed to meet
IEC 801, -2, -3, -4, -5, -6, Level 3
Humidity: Operating; non-condensing
5% to 95%
Vibration: Three orthogonal axes, sweep at 1
oct/min, 5 min. dwell at four major resonances
0.75 G peak 5 Hz to 500 Hz, operational
Storage temperature: -40° to 85°C
Temperature coefficient: ±.04% per °C
MTBF demonstrated: >550,000 hours at full load
and 25°C ambient conditions
Safety
VDE
UL
CSA
CE
0750/EN60601-1 (IEC601)
UL2601
CSA 22.2 No. 601.1
Mark (LVD)
21310-3336-0015
E182560
LR109492
rev 08.11.04
94
Technical Support: (888) 41-ASTEC or (407) 241-2752
Americas: (760) 930-4600
Europe (UK) 44 (1384) 842-211
Asia (HK) 852-2437-9662