Embedded Power for
Business-Critical Continuity
LPS100-M Series
150 Watts
Total Power:
Input Voltage:
# of Outputs:
80 - 150 Watts
90 - 264 Vac
Single
Rev. 1.16.09_114
LPS100-M Series
1 of 3
Special Features
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Medical and ITE safeties
Active power factor correction
2” x 4” footprint
Less than 1U high
EN61000-3-2 compliant
Remote sense
Power fail
Adjustable main output
Built-in Class B EMI filter
Overvoltage protection
Overload protection
Thermal overload protection
Isolated12 V Fan output
LPX50 Enclosure kit available
Electrical Specifications
Input
Input range:
Frequency:
Inrush current:
Efficiency:
EMI/RFI:
Power factor:
Safety ground
leakage current:
90 - 264 Vac; 120 - 300 Vdc
47-63 Hz
50 A max., cold start @ 25 °C
88% typical at full load
FCC Class B conducted; CISPR22 Class B conducted;
EN55022 Class B conducted; VDE0878PT3 Class B conducted
0.99 typical
275 µA @ 50/60 Hz, 264 Vac input
Output
100 W for convection (80 W for LPS102-M);
150 W with 30 CFM forced air (120 W for LPS102-M)
Adjustment range:
±10% minimum on the main outputs
Fan output:
12 V @ 1 A isolated, ±5%
Hold-up time:
10 ms @ 150 W load, 120 Vac input
Overload protection: Short circuit protection on all outputs. Case overload
protected @ 110-160% above rating
Overvoltage protection: 15-35% above nominal output
Maximum power:
Logical Control
Power failure:
Safety
•
•
•
•
•
•
•
TUV
60950, 60601-1
UL
60950, 60601-1
CSA
60950, 60601-1
NEMKO
60950, 60601-1
AUSTEL
60950, 60601-1
CB
Certificate & report
CE
Mark (LVD)
Remote sense:
Open collector logic signal goes high 100-500 msec after main output; it
goes low at least 6 msec before loss of regulation
Compensates for 0.5 V lead drop min. Will operate without remote sense
connected. Reverse connection protected.