Embedded Power for
Business-Critical Continuity
Ordering Information
Output
Voltage
+5 V
+12 V
–12 V
+5 V
+24 V
+5 V
+12 V
+12 V
+15 V
+24 V
Output Current
Max
7A
2.5 A
0.5 A
7A
2A
7A
2.5 A
5.4 A
4.4 A
2.7 A
(1)
Rev.11.2.09_104
NLP65 Medical
3 of 4
Total
Regulation
(6)
±2.0%
±5.0%
±5.0%
±2.0%
±5.0%
±2.0%
±5.0%
±2.0%
±2.0%
±2.0%
NLP65-9912J
NLP65-9915J
NLP65-9924J
NLP65-9929J
NLP65-9920J
Peak
9.1 A
3.3 A
0.81 A
9.1 A
2.6 A
9.1 A
3.3 A
7A
5.7 A
3.5 A
Fan
Ripple
(4)
(10)
Model Number
(11, 12)
NLP65-9908J
8A
3A
1A
8A
2A
8A
3A
6.5 A
5.3 A
3.5 A
50 mV
150 mV
120 mV
50 mV
240 mV
50 mV
150 mV
120 mV
150 mV
240 mV
Notes
Natural convection cooling. Models NLP65-9929J, and NLP65-9908Jmust not exceed 62.5 Watts continuous output power with natural convection. Model
NLP65-9920J not to exceed 65 Watts continuous output power with natural convection.
2
When the input voltage is less than 90 Vac the operating temperature range is 0
°
C to +40
°
C. The ripple and regulation specifications may not be met.
3
Peak output current lasting less than 60 seconds with duty cycle less than 5%. During peak loading, output voltage may exceed total regulation limits.
4
Figure is peak-to-peak for convection power rating. Output noise measurements are made across a 20 MHz bandwidth using a 6 inch twisted pair, terminated
with a 10 µF electrolytic capacitor and a 0.1 µF ceramic capacitor.
5
Three orthogonal axes, random vibration 10 minutes for each axes, 2.4 G rms 5 Hz to 500 Hz.
6
To maintain stated regulation then:
for single output units
I ≥ 0.2 A I max.
for multiple output units
0.25 ≤ I(A)/I(B) ≤ 5, for I(A) ≥ 0.2 A I(A) max.
7
For optimum reliability, no part of the heatsink should exceed 120
o
C, and no semiconductor case temperature should exceed 130
o
C.
8
CAUTION: Allow a minimum of 1 second after disconnecting line power when making thermal measurements.
9
This product is only for inclusion by professional installers within other equipment and must not be operated as a stand alone product.
10
Maximum continuous output power for all multiple output models must not exceed 75 Watts with 20 CFM forced air cooling at 50
o
C.
11
The ‘J’ suffix indicates that these parts are Pb-free (RoHS 6/6) compliant.
12
NOTICE: Some models do not support all options. Please contact your local Emerson Network Power representative or use the on-line model number search
tool at http://www.PowerConversion.com to find a suitable alternative.
1