Pre-amplified/5V Excitation/Gauge
Features
.
.
.
.
Accuracy, 1.0 FS
Volt level output
On-chip amplification and temperature compensations
Pre-calibration of offset voltage and span
XFPM(NIBP) Data sheet
Applications
. Non-invasive Blood Pressure(NIBP)monitors
. Medical device
Part number for ordering
XFPM - 050KP G R - P1
Prefix (NIBP)
P1 : Accuracy 1.0 FS
Model
XFPM
Rated pressure(Pa)
Pressure type
G : Gauge
Terminal leads direction (See Outline Diagram)
No mark :
R
:
Pressure type
Gauge pressure
XFPM
Model
Package configuration
Measurable pressure range (kPa)
0 50
Dual-In-line-Package (DIP)
Part number for ordering
XFPM-050KPG-P1
XFPM-050KPGR-P1
Specifications
Model/Rated pressure
050KPG(R)-P1
Unit
Recommended operating conditions
Pressure type
Gauge pressure
Rated pressure
300(40)
Measurable pressure range
0 300(0 40)
Pressure media
Non-corrosive gas only
Excitation voltage
5 0.25
Absolute maximum rating
Maximum load pressure
Twice of rated pressure
Maximum excitation voltage
8
Operating temperature
40 125
Storage temperature
40 125
Operating humidity
30 80 (No dew condensation)
Electric performances/characteristics (Excitation voltage Vcc=5.0V constant, Ambient temperature Ta=25 )
Current consumption
less than 10
Output impedance
less than 10
Source current
less than 0.2
Sink current
less than 2
Mechanical response time
2 (For the reference)
Full scale span voltage
3.5
Offset voltage 1
0.5 0.1
4.0 0.1
Full scale span voltage 1
Temperature sensitivity of offset(TSO)
0.2
Accuracy 1,2
1.0
Note ; 1
)
Excluding input voltage error.
2
)
Excludes offset calibration error and temperature error of offset.
mmHg(kPa)
mmHg(kPa)
VDC
VDC
RH
mA
mA
mA
msec
V
V
V
V/0 50
FS/0 50
1