ASDX Series Silicon Pressure Sensors
Table 1. Absolute Maximum Ratings
Parameter
Supply voltage (V
supply
)
Voltage to any pin
ESD susceptibility (human body model)
Storage temperature
Lead temperature (2 s to 4 s)
2
External capacitance between V
supply
and ground
Table 2. Operating Specifications
Parameter
3
Supply voltage (V
supply
)
3.3 Vdc
5.0 Vdc
Sensors are either 3.3.Vdc or 5.0.Vdc based on models chosen per the
order guide (see Figure 1).
1
Min.
-0.3
-0.3
3
-50 [-58]
-
100
Max.
6.0
V
supply
+ 0.3
-
125 [257]
250 [482]
470
Unit
Vdc
Vdc
kV
°C [°F]
°C [°F]
nF
Min.
3.0
4.75
1.5
0 [32]
-20 [-4]
Typ.
3.3
4
5.0
4
Max.
3.6
5.25
Unit
Vdc
mA
°C [°F]
°C [°F]
Supply current
5
Compensated temperature range
6
Operating temperature range
7
Overpressure
8
Burst pressure
Startup time (power up to data ready)
Response time
Upper output clipping limit
Lower output clipping limit
Minimum load resistance
9
Total error band
Output resolution
Table 3. Environmental Specifications
Parameter
Humidity
Vibration
Shock
Life
Table 4. Wetted Materials
Parameter
Covers
Adhesives
Electronic components
11
-
-
97.5
-
5.0
-
12
2.5
3.5
-
85 [185]
-
105 [221]
2X operating pressure range minimum
3X operating pressure range minimum
-
5
1.0
-
-
-
-
2.5
-
-
-
2.0
-
-
ms
ms
% V
supply
% V
supply
kOhm
10
%FSS
bits
Characteristic
0% to 95% RH non-condensing
10 G at 20 Hz to 2000 Hz
100 G for 11 ms
1 million cycles minimum
Port 1 (Pressure Port)
glass-filled PBT
silicone
silicon and glass
12
Port 2 (Reference Port)
glass-filled PBT
silicone and epoxy
silicon, glass, and gold
12
Notes:
1.
Absolute maximum ratings are the extreme limits that the device will withstand without damage to the device.
2.
An external bypass capacitor is
required
across the supply voltage (Pins 1 and 3 – see Figure 4) as close to the sensor supply pin as possible for correct sensor
operation.
3.
Ratiometricity of the sensor (the ability of the output to scale to the input voltage) is achieved within the specified operating voltage for each option. Other custom
supply voltages are available, please contact Honeywell Customer Service.
4.
The sensor is not reverse polarity protected. Incorrect application of excitation voltage or ground to the wrong pin may cause electrical failure.
5.
The compensated temperature range is the temperature range (or ranges) over which the sensor will produce an output proportional to pressure within the
specified performance limits.
6.
The operating temperature range is the temperature range over which the sensor will produce an output proportional to pressure but may not remain within the
specified performance limits.
7.
Overpressure is the maximum pressure which may safely be applied to the product for it to remain in specification once pressure is returned to the operating
pressure range. Exposure to higher pressures may cause permanent damage to the product.
8.
Burst pressure is the maximum pressure that may be applied to any port of the product without causing escape of pressure media. Product should not be expected
to function after exposure to any pressure beyond the burst pressure.
9.
Total error band is the maximum deviation in output from ideal transfer function over the entire compensated temperature and pressure range. Includes all errors
due to offset, full scale span, pressure non-linearity, pressure hysteresis, repeatability, thermal effect on offset, thermal effect on span and thermal hysteresis.
Specification units are in percent of full scale span (%FSS).
10. Full scale span (FSS) is the algebraic difference between the output signal measured at the maximum (Pmax.) and minimum (Pmin.) limits of the pressure range.
11. Consult Honeywell Customer Service for detailed material information.
12. For the AC pressure port configuration, the “pressure” and “reference” ports are reversed.
2