欢迎访问ic37.com |
会员登录 免费注册
发布采购

HTB400-P/SP5 参数 Datasheet PDF下载

HTB400-P/SP5图片预览
型号: HTB400-P/SP5
PDF下载: 下载PDF文件 查看货源
内容描述: 电流转换器HTB 50 〜 400 -P / SP5和HTB 50 〜 100 -TP / SP5 [Current Transducers HTB 50~400-P/SP5 and HTB 50~100-TP/SP5]
分类和应用: 转换器
文件页数/大小: 2 页 / 66 K
品牌: LEM [ LEM ]
 浏览型号HTB400-P/SP5的Datasheet PDF文件第2页  
Current Transducers HTB 50 .. 400-P/SP5
and HTB 50 .. 100-TP/SP5
For the electronic measurement of currents: DC, AC, pulsed, mixed,
with a galvanic isolation between the primary circuit (high power) and
the secondary circuit (electronic circuit).
I
PN
= 50 .. 400 A
Electrical data
Primary nominal
r.m.s. current
I
PN
(A)
50
100
200
300
400
Primary current
measuring range
I
P
(A)
±150
±300
±500
±600
±600
Type
HTB 50-P/SP5, HTB 50-TP/SP5
1)
HTB 100-P/SP5, HTB 100-TP/SP5
1)
HTB 200-P/SP5
HTB 300-P/SP5
HTB 400-P/SP5
Features
Hall effect measuring principle
Galvanic isolation between primary
and secondary circuit
Isolation voltage 2500V
Low power consumption
Primary bus bar option for 50A and
100A version for ease of connection
V
C
I
C
V
d
R
IS
V
OUT
R
OUT
R
L
Supply voltage (±5 %)
2)
Current consumption
R.m.s. voltage for AC isolation test, 50/60 Hz, 1 mn
Isolation resistance @ 500 VDC
Output voltage @ ±
I
PN
,
R
L
= 10 kΩ,
T
A
= 25°C
Output internal resistance
Load resistance
+12 .. +15
V
<15
mA
2.5
kV
>500
MΩ
V
OE
±1.667
V
100
≥10
kΩ
Accuracy - Dynamic performance data
Special Features
Single power supply from 12V to 15V
ε
X
L
V
OE
V
OH
V
OT
TC
ε
G
t
r
f
Accuracy @
I
PN
,
T
A
= 25°C (without offset)
<±1
% of
I
PN
Linearity (0 .. ±
I
PN
)
<±1
% of
I
PN
Electrical offset voltage,
T
A
= 25°C
Vc/2±30
mV
Hysteresis offset voltage @
I
P
= 0;
after an excursion of 3 x
I
PN
<±0.5 % of
I
PN
Thermal drift of
V
OE
HTB 50-(T)P/SP5
<±1.0
mV/K
HTB 100-(T)P..400-P/SP5
<±0.5
mV/K
Thermal drift (% of reading)
<±0.05
%/K
Response time @ 90% of
I
P
<3
µs
3)
Frequency bandwidth (-3 dB)
DC .. 50 kHz
Advantages
Small size and space saving
Only one design for wide current
ratings range
High immunity to external
interference.
General data
T
A
T
S
m
Ambient operating temperature
Ambient storage temperature
Mass (-TP version)
-25 .. +85
-25 .. +85
<30 (<36)
°C
°C
g
Applications
AC variable speed drives
Static converters for DC motor drives
Battery supplied applications
Uninterruptible Power Supplies
(UPS)
Switched Mode Power Supplies
(SMPS)
Power supplies for welding
applications.
Notes : EN 50178 approval pending
1)
-TP version is equipped with a primary bus bar.
2)
Operating at +12V
Vc < +15V will reduce measuring range.
3)
Derating is needed to avoid excessive core heating at high frequency.
030221/0
LEM Components
www.lem.com