RPS-60-12 参数 Datasheet PDF下载

RPS-60-12图片预览
型号: RPS-60-12
PDF下载: 下载PDF文件 查看货源
内容描述: 60W单输出医用型 [60W Single Output Medical Type]
分类和应用:
文件页数/大小: 2 页 / 287 K
品牌: MEANWELL [ MEAN WELL ENTERPRISES CO., LTD. ]
 浏览型号RPS-60-12的Datasheet PDF文件第2页  
60W Single Output Medical Type
Features :
4"x2" miniature size
Universal AC input/Full range
Low leakage current<200uA
RPS-60
series
Protections: Short circuit / Overload / Over voltage
Cooling by free air convection
UL60601-1/IEC60601-1/EN60601-1 medical safety approved
UL60950-1/IEC60950-1/EN60950-1 ITE safety approved
No load power consumption<0.75W
Fixed switch frequency at 100KHz
3 years warranty
EN60601-1
IEC60601-1
SPECIFICATION
MODEL
DC VOLTAGE
RATED CURRENT
CURRENT RANGE
RATED POWER
PEAK LOAD(10sec.)
OUTPUT
VOLTAGE ADJ. RANGE
VOLTAGE TOLERANCE
Note.3
LINE REGULATION
LOAD REGULATION
SETUP, RISE TIME
HOLD UP TIME (Typ.)
VOLTAGE RANGE
FREQUENCY RANGE
INPUT
EFFICIENCY (Typ.)
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
LEAKAGE CURRENT
OVER LOAD
PROTECTION
OVER VOLTAGE
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
WITHSTAND VOLTAGE
RPS-60-3.3
3.3V
10A
0 ~ 11A
33W
Note.4
36.3W
RPS-60-5
5V
10A
0 ~ 11A
50W
55W
80mVp-p
4.75 ~ 5.5V
2.0%
0.5%
1.0%
RPS-60-12
12V
5A
0 ~ 5.5A
60W
66W
120mVp-p
11.4 ~ 13.2V
2.0%
0.5%
1.0%
RPS-60-15
15V
4A
0 ~ 4.4A
60W
66W
150mVp-p
13.5 ~ 16.5V
2.0%
0.5%
1.0%
RPS-60-24
24V
2.5A
0 ~ 2.75A
60W
66W
240mVp-p
22.8 ~ 27.6V
1.0%
0.5%
1.0%
RPS-60-48
48V
1.25A
0 ~ 1.375A
60W
66W
300mVp-p
45.6 ~ 52.8V
1.0%
0.5%
1.0%
RIPPLE & NOISE (max.)
Note.2
80mVp-p
3.1 ~ 3.6V
2.0%
0.5%
1.0%
500ms, 30ms/230VAC
50ms/230VAC
90 ~ 264VAC
47 ~ 63Hz
74%
1.8A/115VAC
79%
500ms, 30ms/115VAC at full load
13ms/115VAC at full load
127 ~ 370VDC
83%
30A/115VAC
84%
85%
86%
1 A/230VAC
COLD START 60A/230VAC
For earth <200uA / 264VAC, For patient <100uA/264VAC
115 ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
3.8 ~ 5V
5.75 ~ 6.75V
13.8 ~ 16.2V
17.25 ~ 20.25V
Protection type : Shut down o/p voltage, re-power on to recover
-20 ~ +70
(Refer to output load derating curve)
20 ~ 90% RH non-condensing
-40 ~ +85 , 10 ~ 95% RH
0.03%/ (0 ~ 45 )
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
UL60950-1, TUV EN60950-1, UL60601-1, TUV EN60601-1, IEC60601-1 approved
28.4 ~ 32.4V
55.2 ~ 64.8V
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:1.5KVAC
SAFETY & ISOLATION RESISTANCE
I/P-O/P, I/P-FG, O/P-FG:100M Ohms/500VDC
EMC
EMI CONDUCTION & RADIATION
Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B
(Note 5)
HARMONIC CURRENT
EMS IMMUNITY
MTBF
OTHERS
DIMENSION
PACKING
NOTE
Compliance to EN61000-3-2,-3
Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A
353.6Khrs min. MIL-HDBK-217F (25
101.6*50.8*29mm (L*W*H)
0.15Kg; 96pcs/15.4Kg/0.89CUFT
)
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.
5. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives.
6. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
File Name:RPS-60-SPEC 2008-03-25