R Series Intelligent DAQ –
Data Acquisition and Control with Onboard Processing
Environmental
NI 78xxR devices are intended for indoor use only.
Safety and Compliance
Safety
NI 78xxR devices are designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 61010-1, CSA 61010-1
Note:
For UL and other safety certifications, refer to the product label or
visit
ni.com/certification,
search by model number or product line, and
click the appropriate link in the Certification column.
Operating Environment
NI 781xR.............................................. 0 to 55 °C (tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2)
NI 7830R, NI 7831R
40 or 80 MHz timebase .................. 0 to 55 °C (tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2)
NI 7833R/7841R/7842R/7851R/7852R/7853R/7854R
40 MHz timebase ........................... 0 to 55 °C (tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2)
80 MHz timebase ........................... 0 to 55 °C except the
following: 0 to 45 °C when
installed in an NI PXI-1000/B or
NI PXI-101x (tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2)
Relative humidity range...................... 10 to 90%, noncondensing
(tested in accordance with
IEC-60068-2-56)
Altitude................................................ 2,000 m at 25 °C
ambient temperature
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards
for electrical equipment for measurement, control, and laboratory use:
• EN 61326 (IEC 61326): Class A emissions; Basic immunity
• EN 55011 (CISPR 11): Group 1, Class A emissions
• AS/NZS CISPR 11: Group 1, Class A emissions
• FCC 47 CFR Part 15B: Class A emissions
• ICES-001: Class A emissions
CE Compliance
This product meets the essential requirements of applicable European
Directives, as amended for CE marking, as follows:
• 2006/95/EC; Low-Voltage Directive (safety)
• 2004/108/EC; Electromagnetic Compatibility Directive (EMC)
Note:
Refer to the Declaration of Conformity (DoC) for this product for
any additional regulatory compliance information. To obtain the DoC for
this product, visit
ni.com/certification,
search by model number or
product line, and click the appropriate link in the Certification column.
Storage Environment
Ambient temperature range ............... -20 to 70 °C (tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2)
Relative humidity range...................... 5 to 95%, noncondensing
(tested in accordance with
IEC-60068-2-56)
Note:
Clean the device with a soft, nonmetallic brush. Make sure that
the device is completely dry and free from contaminants before returning
it to service.
Environmental Management
National Instruments is committed to designing and manufacturing
products in an environmentally responsible manner. NI recognizes that
eliminating certain hazardous substances from our products is beneficial
not only to the environment but also to NI customers.
For additional environmental information, refer to the NI and the
Environment Web page at
ni.com/environment.
This page contains the
environmental regulations and directives with which NI complies, as
well as other environmental information not included in this document.
Shock and Vibration (for NI PXI-78xxR Only)
Operational shock ............................... 30 g peak, half-sine, 11 ms
pulse (tested in accordance
with IEC-60068-2-27; test
profile developed in accordance
with MIL-PRF-28800F)
Random vibration
Operating ........................................ 5 to 500 Hz, 0.3 g
rms
Nonoperating.................................. 5 to 500 Hz, 2.4 g
rms
(tested in accordance with
IEC-60068-2-64; nonoperating test
profile exceeds the requirements
of MIL-PRF-28800F, Class 3)
Waste Electrical and Electronic Equipment (WEEE)
EU Customers:
At the end of their life cycle, all products must be
sent to a WEEE recycling center. For more information about WEEE
recycling centers and National Instruments WEEE initiatives,
visit
ni.com/environment/weee.htm.
BUY ONLINE
at
ni.com
or
CALL 800 813 3693 (U.S.)
7