欢迎访问ic37.com |
会员登录 免费注册
发布采购

NIPXI-5404 参数 Datasheet PDF下载

NIPXI-5404图片预览
型号: NIPXI-5404
PDF下载: 下载PDF文件 查看货源
内容描述: 9 kHz至105 MHz的正弦波/ DC至100 MHz的时钟/ 1.07 μHz的频率分辨率 [9 kHz to 105 MHz sine wave/DC to 100 MHz clock/1.07 μHz frequency resolution]
分类和应用: 时钟
文件页数/大小: 6 页 / 786 K
品牌: NI [ NATIONAL INSTRUMENTS CORPORATION ]
 浏览型号NIPXI-5404的Datasheet PDF文件第2页浏览型号NIPXI-5404的Datasheet PDF文件第3页浏览型号NIPXI-5404的Datasheet PDF文件第4页浏览型号NIPXI-5404的Datasheet PDF文件第5页浏览型号NIPXI-5404的Datasheet PDF文件第6页  
100 MHz Frequency Generator
NI PXI-5404
• 1 channel updated at 300 MS/s
• 9 kHz to 105 MHz sine wave
• DC to 100 MHz clock
• 1.07 µHz frequency resolution
• 1 to 4 V
pp
sine amplitude
• 12-bit vertical resolution
• ±0.2 dB flatness of sine wave
passband (9 kHz to 100 MHz)
• Ideal for coherent sampling
applications
• Phase-lock loop for synchronization
to other devices and PXI backplane
New
Operating Systems
• Windows 2000/NT/XP/Me/9x
Recommended Software
• LabVIEW
• LabWindows/CVI
Other Compatible Software
• Visual Basic
• C/C++
Driver Software
(included)
• NI-FGEN
Calibration Certificate Included
Overview
The National Instruments PXI-5404 is a 100 MHz frequency
generator packaged in 1-slot PXI module. It generates sine and clock
outputs simultaneously at up to 100 MHz with 1.07 µHz resolution.
By using direct digital synthesis (DDS) and precise phase-lock loop
circuitry, the PXI-5404 generates stable, accurate sine and clock
outputs of which you can programmatically adjust the amplitude,
frequency, and phase of the sine and clock outputs as well as the
duty cycle of the clock output.
The PXI-5404 is a versatile, cost-effective, sine wave generator
with output levels ranging from 1 to 2
Vpp
into a 50
load with
12-bit resolution, excellent frequency resolution, and passband
flatness. Stimulus-response instruments such as network analyzers
and ATE test systems will benefit from using the PXI-5404 as the
sine stimulus because of its passband flatness, and its frequency
accuracy and resolution.
The PXI-5404 can also serve as a programmable clock generator
for ATE systems because of its 1.07 µHz frequency resolution, duty-
cycle adjustability from 25 to 75 percent, and programmable voltage
levels of 5, 3.3, or 1.8 V into high-impedance loads.
100100...
111001...
101001...
ADC
(DUT)
Ideal for Coherent
Clocking Applications
With the simultaneous sine and clock
outputs, you can employ coherent
clocking to reduce the test time of your
device under test (DUT) and eliminate
the need for antialias filtering. Because
you have a precise phase and
frequency relationship between the
INFO CODES
For more information
or to order products
online, visit
ni.com/info
and enter:
pxi5404
BUY ONLINE!
analog stimulus and the sample clock, there is no need for window-
weighting functions in your FFTs for spectral analysis, thus reducing
the need to sample the DUT over several cycles of the sine stimulus.
With a coherent clocking system, you know both the analog
stimulus and the sampling clock frequencies, thus eliminating the
need for antialias filters or time-domain window functions. In a
coherent sampling system, all the power from the signal and its
accompanying harmonics fall into predictable frequency bins.
NI PXI-5404
Sine
Clock
PLL
Figure 1. A typical
coherent sampling
application is the testing
of analog-to-digital
converters (ADCs). Here
the sine stimulus is fed to
the ADC and the clock
output is phase locked
and multiplied to yield a
sampling clock to the ADC
and to a digital pattern
generator/analyzer.
Digital I/O module