欢迎访问ic37.com |
会员登录 免费注册
发布采购

NIPXI-4060 参数 Datasheet PDF下载

NIPXI-4060图片预览
型号: NIPXI-4060
PDF下载: 下载PDF文件 查看货源
内容描述: 51 ? 2位数字万用表 [51?2 Digit Multimeters]
分类和应用:
文件页数/大小: 4 页 / 110 K
品牌: NI [ NATIONAL INSTRUMENTS CORPORATION ]
 浏览型号NIPXI-4060的Datasheet PDF文件第1页浏览型号NIPXI-4060的Datasheet PDF文件第2页浏览型号NIPXI-4060的Datasheet PDF文件第4页  
5
1
2
Digit Multimeters
5
1
2
Digit Multimeters
Specifications – NI 4060
Specifications are guaranteed between 15 and 35 °C unless otherwise noted.
AC Current
(Accuracy % of reading ± mA)
Range
20.0000 mA
200.000 mA
10.0000 A*
24 Hour
(25 °C ± 1 °C)
0.30% + 100 µA
0.18% + 0.22 mA
0.3% + 22 mA
90 Day
(25 °C ± 10 °C)
0.32% + 170 µA
0.20% + 1.2 mA
0.32% + 120 mA
1 Year
Temperature Coefficient
(25 °C ± 10 °C) (% of reading/°C ± mA/°C)
0.32% + 170 µA 0.022%/°C + 12 µA/°C
0.20% + 1.2 mA 0.009%/°C + 0.110 mA/°C
0.32% + 120 mA 0.026%/°C + 11 mA/°C
DC Voltage
(Accuracy % of reading ± µV)
Range
250.000 V
25.0000 V
2.00000 V
200.000 mV
20.0000 mV
24 Hour
(25 °C ± 1 °C)
0.0032% + 1.25 mV
0.0032% + 1 mV
0.0029% + 10 µV
0.0029% + 6 µV
0.0029% + 6 µV
90 Day
(25 °C ± 10 °C)
0.021% + 1.25 mV
0.021% + 1 mV
0.014% + 10 µV
0.014% + 6 µV
0.014% + 6 µV
1 Year
(25 °C ± 10 °C)
0.024% + 1.25 mV
0.024% + 1 mV
0.017% + 10 µV
0.017% + 6 µV
0.017% + 6 µV
Temperature Coefficient
(% of reading/°C + µV/°C)
0.0017%/°C + 480 µV/°C
0.0017%/°C + 480 µV/°C
0.0009%/°C + 5 µV/°C
0.0009%/°C + 1 µV/°C
0.0009%/°C + 1 µV/°C
*Requires 10 A shunt, CSM-10A
Accuracy numbers are for 5
1
2
digits and include the effects of full and zero-scale
errors, temperature variation, linearity, and noise.
Accuracy numbers are for 5
1
2
digits with autozero on and include the effects of
full and zero-scale errors, temperature variation, linearity, and noise. Overrange
capabilities to 300 V.
Input Characteristics
Input protection ......................................... Fuse F1 500 mA/250 V fast-fusing
Shunt resistor ............................................ 1
Burden voltage........................................... < 400 mV at 200 mA AC
Noise Rejection
NMRR ........................................................ 80 dB (10 Hz filter setting, 50/60 Hz
powerline frequency ±1%)
DC ECMRR ................................................ 140 dB (with a 1 kΩ imbalance in HI lead)
AC ECMR (50/60 Hz) ................................. 150 dB (with a 1 kΩ imbalance in HI lead)
Resistance
(Accuracy % of reading ±
Ω)
Range
Extended ohm
(> 2 MΩ)
2.00000 MΩ*
200.000 kΩ
20.0000 kΩ
2.00000 kΩ
200.000
24 Hour
(25 °C ± 1 °C)
0.1% + 6 kΩ
0.012% + 9
0.012% + 5
0.006% + 0.09
0.006% + 0.05
0.006% + 0.05
90 Day
(25 °C ± 10 °C)
0.1% + 60 kΩ
0.077% + 27
0.077% + 22
0.024% + 0.3
0.024% + 0.2
0.024% + 0.2
1 Year
Temperature Coefficient
(25 °C ± 10 °C) (% of reading/°C +
Ω/°C)
0.1% + 60 kΩ
0.080% + 27
0.080% + 22
0.027% + 0.3
0.027% + 0.2
0.027% + 0.2
0.0072%/°C + 6 kΩ/°C
0.0072%/°C + 2
Ω/°C
0.0072%/°C + 2
Ω/°C
0.0020%/°C + 0.02
Ω/°C
0.0020%/°C + 0.02
Ω/°C
0.0020%/°C + 0.02
Ω/°C
Input Characteristics
Input bias current....................................... 1 nA maximum
Input resistance ......................................... >1 GΩ (2 V, 200 mV, 20 mV ranges);
1 MΩ (250 V, 25 V)
DC Current
(Accuracy % of reading ± µA)
Range
20.0000 mA
200.000 mA
10.0000 A*
24 Hour
(25 °C ± 1 °C)
0.015% + 10 µA
0.015% + 10 µA
0.02% + 1 mA
90 Day
(25 °C ± 10 °C)
0.039% + 10 µA
0.039% + 10 µA
0.035% + 2 mA
1 Year
Temperature Coefficient
(25 °C ± 10 °C) (% of reading/°C + µA/°C)
0.042% + 10 µA
0.0035%/°C + 1 µA/°C
0.042% + 10 µA
0.0035%/°C + 1 µA/°C
0.035% + 2 mA
0.007%/°C + 0.1 mA/°C
*Requires 10 A shunt, CSM-10A
Accuracy numbers are for 5
1
2
digits with autozero on and include the effects of
full and zero-scale errors, temperature variation, linearity, and noise.
*With autozero on, or while scanning, and when large resistance with capacitive
loads is measured, additional delay time is required.
Accuracy numbers are for 4-wire resistance mode, 5
1
2
digits with autozero on and
include the effects of full and zero-scale errors, temperature variation, linearity, and noise.
Measurement modes
Ohms.................................................... 2 or 4-wire
Extended ohms .................................... 2-wire only
Maximum lead resistance.......................... 10
(200
range), 1 kΩ
(all other ranges)
Additional error for 2-wire resistance ........ 0.6
Input Characteristics
Input protection ......................................... Fuse F1 500 mA/250 V fast-fusing
Shunt resistor ............................................ 1
Burden voltage........................................... <400 mV at 200 mA DC
Diode Testing
(Accuracy % of reading ± µV)
Range
2V
24 Hour
(25 °C ± 1 °C)
0.006% + 7 µV
90 Day
(25 °C ± 10 °C)
0.024% + 22 µV
1 Year
Temperature Coefficient
(25 °C ± 10 °C) (% of reading/°C + µV/°C)
0.027% + 22 µV
0.0020%/°C + 2 µV/°C
Modular Instrumentation
AC Voltage
(Accuracy % of reading ± mV)
Range
250.000 V
25.0000 V
2.00000 V
200.000 mV
20.0000 mV
24 Hour
(25 °C ± 1 °C)
0.6% + 250 mV
0.16% + 30 mV
0.28% + 3 mV
0.16% + 0.22 mV
0.28% + 100 µV
90 Day
(25 °C ± 10° C)
0.62% + 680 mV
0.18% + 210 mV
0.30% + 21 mV
0.18% + 1.20 mV
0.30% + 170 µV
1 Year
Temperature Coefficient
(25 °C ± 10 °C) (% of reading/°C + mV/°C)
0.62% + 680 mV
0.007%/°C + 20 mV/°C
0.18% + 210 mV 0.007%/°C + 20 mV/°C
0.30% + 21 mV
0.019%/°C + 2 mV/°C
0.18% + 1.20 mV 0.007%/°C + 0.110 mV/°C
0.30% + 170 µV
0.019%/°C + 12 µV/°C
Accuracy numbers are for 5
1
2
digits with autozero on and include the effects of
full and zero-scale errors, temperature variation, linearity, and noise.
Test current................................................ 100 µA
General Specifications
Settling time .............................................. Affected by source impedance and
input signal changes.
Warm-up time ............................................ 30 minutes for measurements
accurate within specifications.
Altitude....................................................... Up to 2,000 m; at higher altitudes the
installation category must be derated.
Working voltage ......................................... 300 V maximum between either input
terminal and earth ground
Accuracy numbers are for 5
1
2
digits and include the effects of full and zero-scale
errors, temperature variation, linearity, and noise. Applies for sine waves
10% of
input range. Accuracy may be affected by source impedance, cable capacitances,
dielectric absorption, or slew rate. Overrange capabilities to 300 V.
Noise Rejection
AC CMRR (50/60 Hz) ................................ >80 dB (with a 1 kΩ imbalance in HI lead)
Power Requirement
+5 VDC ......................................................
250 mA in operational mode
Input Characteristics
Input resistance ......................................... 1 MΩ
Bandwidth.................................................. 20 Hz - 25 kHz
Physical
Dimensions
PCI ........................................................ 10.8 by 17.5 cm (4.25 by 6.9 in.)
PXI ........................................................ 10 by 16 cm (3.9 by 6.33 in.)
Additional AC Errors
Frequency dependent errors
Input Frequency Additional Error (% of reading)
20 Hz - 50 Hz
2.5%
50 Hz - 100 Hz
0%
100 Hz - 20 kHz
1%
20 Hz - 25 kHz
2.5%
Environment
Operating temperature .............................. 0 to 55 °C
Storage temperature.................................. -20 to 70 °C
Relative humidity ....................................... 10 to 90%, noncondensing
Certifications and Compliances
CE Mark Compliance
National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • info@ni.com • ni.com
465