High-Density Multiconfiguration Multiplexer/Matrix
Specifications
PXI-2530
Multiplexer sizes ....................................... 128x1 (1-wire)
64x1 (2-wire)
32x1 (4-wire)
16x1 (1-wire) (8)
Matrix sizes ............................................... 4x32 (1-wire)
8x16 (1-wire)
4x16 (2-wire)
SCXI-1130
Multiplexer sizes ....................................... 256x1 (1-wire)
128x1 (2-wire)
64x1 (4-wire)
16x1 (1-wire) (16)
Matrix sizes ............................................... 4x64 (1-wire)
8x32 (1-wire)
4x32 (2-wire)
Physical
Relay types .......................................................
Contact material...............................................
I/O connectors ..................................................
Dimensions
PXI-2530.....................................................
SCXI-1130 ..................................................
reed
rhodium
176-pin docking station plug
2 x 10 x 16 cm (.8 x 3.9 x 6.4 in.)
3.0 x 17.2 x 20.3 cm (1.2 x 6.9 x 8.0 in.)
Environment
Operating temperature.....................................
Storage temperature ........................................
Relative humidity .............................................
Pollution degree ...............................................
Indoor use only
0 to 55 ˚C
-40 to 70 ˚C
5 to 85% noncondensing
2
Safety
This product is designed to meet the requirements of the following standards of safety for electrical
equipment for measurement, control, and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 3111-1, UL 61010B-1
• CAN/CSA C22.2 No. 1010.1
CE Compliance
This product meets the essential requirements of applicable European Directives,
as amended for CE marking, as follows:
Low-Voltage Directive (safety) .................. 73/23/EEC
Electromagnetic Compatibility
Directive (EMC) ................................... 89/336/EEC
Input Characteristics
Maximum switching voltage............................ 60 VDC, 30 V
rms
, CAT I (channel-to-channel
and channel-to-ground)
Maximum current ............................................. 0.4 A (per terminal or internal path)
Maximum switching power ............................. 10 W (per channel, resistive)
Typical DC path resistance
(channel-to-common)
Initial ................................................... <2
Ω
End of life............................................
≥3 Ω
Dynamic Characteristics
Maximum scan rate ......................................... 900 channels/s
Expected relay life
Mechanical ................................................ 10
9
operations
Electrical .................................................... 10
8
operations (10 V, 100 mA, resistive)
National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • info@ni.com • ni.com