欢迎访问ic37.com |
会员登录 免费注册
发布采购

SCXI-1521 参数 Datasheet PDF下载

SCXI-1521图片预览
型号: SCXI-1521
PDF下载: 下载PDF文件 查看货源
内容描述: 24 1/4桥应变输入/可编程激励 [24 quarter-bridge strain gage inputs/Programmable excitation]
分类和应用:
文件页数/大小: 3 页 / 581 K
品牌: NI [ NATIONAL INSTRUMENTS CORPORATION ]
 浏览型号SCXI-1521的Datasheet PDF文件第2页浏览型号SCXI-1521的Datasheet PDF文件第3页  
SCXI Quarter-Bridge Strain Gage Input Module
NI SCXI-1521, NI SCXI-1521B
• 24 quarter-bridge strain gage inputs
1
350
(SCXI-1521)
120
(SCXI-1521B)
• Programmable excitation
(0 to 5 V) per channel
• Programmable shunt and null
calibration per channel
• 10 Hz filter per channel
• Onboard quarter-bridge completion
and shunt calibration resistors
• NI-DAQmx 7.3 simplifies
configuration and measurements
NEW
Operating Systems
• Windows 2000/NT/XP
Recommended Software
• LabVIEW
• LabWindows/CVI
• Measurement Studio
Other Compatible Software
• Visual Basic, C, C++
Measurement Services
Software (included)
• NI-DAQmx 7.3
Calibration Certificate Included
half-bridge or full-bridge strain gage
measurements, please consider the SCXI-1520.
1
For
Overview
The National Instruments SCXI-1521 and SCXI-1521B are quarter-
bridge strain gage input modules that offer all of the features you need
for high-channel-count applications. Each NI SCXI-1521/B is shipped
with a NIST-traceable calibration certificate.
Each SCXI-1521/B module can multiplex its signals into a single
channel of the controlling DAQ device, and you can add more
modules to increase channel count.
Null Compensation
Each input channel of an SCXI-1521/B includes a circuit to remove
bridge offset voltage. Driver software nulls the offset voltage to zero
in seconds. You do not need to manually adjust a potentiometer. By
removing this offset through the measurement hardware, you can
achieve better measurement resolution.
Shunt Calibration
Each input channel of an SCXI-1521/B includes an accurate and
stable shunt calibration resistor with which you can progamatically
simulate loading effects on your strain gage to compensate for
strain gage errors. The shunt calibration resistors located inside of
the module. You enable or disable the shunt resistors through
software commands.
Analog Input
Each of the 24 analog inputs of the SCXI-1521/B consists of a
quarter-bridge completion resistor (350
for the SCXI-1521 and
120
for the SCXI-1521B), and a 10 Hz, 2-pole Butterworth filter.
You can program the excitation over 1024 different levels from 0 to
5 V. The 2-pole Butterworth filters provides lowpass filtering to block
noise and ensure measurement accuracy.
Measurement Services Software
NI-DAQmx is the robust measurement services software included with all
National Instruments data acquisition and signal conditioning products.
This easy-to-use software tightly integrates the full functionality of your
DAQ hardware to LabVIEW, LabWindows/CVI, and Measurement Studio
for Visual Basic. High-performance features include multidevice
synchronization, networked measurements, and DMA data management.
Bundled with NI-DAQmx, the Measurement & Automation Explorer
(MAX) utility simplifies the configuration of your measurement hardware
with device test panels, interactive measurements, and scaled I/O channels.
NI-DAQmx also provides numerous example programs for LabVIEW
and other application development environments to get you started with
your application quickly.
Special Functions
Part Number
777687-17
Calibration
The SCXI-1521/B modules provide simple yet powerful calibration
capabilities. Each module has an onboard calibration EEPROM that
stores calibration constants for each channel; factory calibration
constants are stored in a protected area of the EEPROM.
NI-DAQ traditional and NI-DAQmx transparently use the
calibration constants to correct for gain and offset errors for
each channel.
Terminal Block
SCXI-1317
Type
Screw terminals
CJ Sensor
Compatible Modules
SCXI-1521 & SCXI-1521B
Front-mounting