欢迎访问ic37.com |
会员登录 免费注册
发布采购

PE125BUV 参数 Datasheet PDF下载

PE125BUV图片预览
型号: PE125BUV
PDF下载: 下载PDF文件 查看货源
内容描述: CERMAX氙弧灯 [CERMAX XENON ARC LAMPS]
分类和应用:
文件页数/大小: 2 页 / 67 K
品牌: PERKINELMER [ PERKINELMER OPTOELECTRONICS ]
 浏览型号PE125BUV的Datasheet PDF文件第1页  
PE125BF and PE125BUV
Operational Specifications
Description
Power
Current
Operating Voltage
Ignition Voltage
Temperature
Lifetime*
Nominal
125 watts
12 amps (DC)
11 volts (DC)
150ºÞ C (maximum)
1000 hours (500 hour warranty)
Range
75-150 watts
7-14 amps (DC)
9.5-12.5 volts (DC)
Physical Specifications
Description
Arc Gap
Reflector Geometry
Weight
Window Diameter
S
pecification
.044 inch (1.14mm)
Parabolic Y
2
= 0.5 X (inch)
130.8 grams
1 inch (25.4 mm)
17 kilovolts (recommended minimum)
Focused Output at F/1.0 Lens
Description
6mm aperture
8mm aperture
Visible Output
1050 Lumens
620 Lumens
Total Output*
9.5 watts
5.6 watts
* End of lamp life is defined as 50% of initial output.
Output at Nominal Power
F= UV Filtered Output/ UV = UV Enhanced Output
Description
Peak Intensity
Radiant Output*
UV Output*
IR Output*
Visible Output*
Color Temperature
Peak Instabilities
Beam Geometry**
PE125BF
300x10
3
candelas
17 watts
0.8 watts
10 watts
1500 Lumens
5600 Kelvin
4%
4.5º/5º/6º
PE125BUV
280x10
3
candelas
17 watts
2.2 watts
9 watts
1350 Lumens
5050 Kelvin
4%
4.5º/5º/6º
* Nominal values at 125 watts after 2 hour burn-in.
Notes
1.
2.
3.
4.
Lamp must not be operated with window facing upwards within
45º of vertical.
Seal temperature must not exceed 150º C.
Current/power regulated power supplies and PerkinElmer lamp
housing units are recommended.
Lamp must be operated within recommended current and power
range. Over powering may lead to arc instability, hard starting
and premature aging.
Hot mirror assembly is available for IR filtering.
CERMAX lamps are much safer lamps to use than their quartz
xenon arc lamp equivalents. However, caution must be practiced
when operating lamps because they are under high pressure,
require high voltage, reach temperatures up to 200º C, and their
IR and UV radiation can cause skin burns and eye damage.
Read hazard sheet included with each lamp shipment.
*These values indicate total output in all directions.
Wavelengths = UV<390nm, IR>770nm, Visible 390nm to 770nm.
** Beam Geometry defined as half angle at 10% PTS after 01/100/1000 hours.
5.
6.
Dimensions
.250 (6.35)
1.000 (25.4)
.250 (6.35)
.164
(4.19)
1.00
(25.4)
Spectral Output
100
F
UV
F + HOT MIRROR
1.667
42.34
.125 (3.18)
Y
X
.035
(.899)
.066 (1.52)
RELATIVE OUTPUT
6-32 UNC 2B x .156 DP
3 Holes EQ SP on A
.750 (19.05) BC
2 x .40
(10.2)
80
1.298 (32.96)
60
40
20
Focal Point
Cathode
0
200
300
400
500
600
700
800
900
1000
1100
WAVELENGTH (NM)
PerkinElmer welcomes inquiries about special lamps. We would be pleased to discuss the requirements of your application and the feasibility of
designing a lamp specifically suited to your needs.
For more information e-mail us at opto@perkinelmer.com or visit our web site at www.perkinelmer.com/opto.
All values are nominal; specifications subject to change without notice.
USA:
PerkinElmer Optoelectronics
44370 Christy St.
Fremont, CA 94538
Phone: (510) 979-6500
Fax: (510) 687-1140
Toll Free: (800) 775-6786
Europe:
PerkinElmer Optoelectronics GmbH
Wenzel-Jaksch-Str. 31
65199 Wiesbaden
Germany
Phone: +49 611 492 0
Fax: +49 611 492 369
Asia:
PerkinElmer Optoelectronics
47 Ayer Rajah Crescent #06-12
Singapore 139947
Phone: +65 775 2022
Fax: +65 775 1008
.
DS-152 Rev C 1101
© 2000 PerkinElmer inc. All rights reserved.
www.perkinelmer.com/opto