TR3005
Designed for Short-Range Wireless Medical Data Communications
Supports RF Data Transmission Rates Up to 115.2 kbps
3 V, Low Current Operation plus Sleep Mode
The TR3005 ultra-miniature hybrid transceiver is ideal for short-range wireless medical data ap-
plications where robust operation, small size, low power consumption and low cost are required.
The TR3005 employs RFM’s amplifier-sequenced hybrid (ASH) architecture to achieve this
unique blend of characteristics. All critical RF functions are contained in the hybrid, simplifying
and speeding design-in. The receiver section of the TR3005 is sensitive and stable. A wide dy-
namic range log detector, in combination with digital AGC and a compound data slicer, provide
robust performance in the presence of on-channel interference or noise. Two stages of SAW fil -
tering provide excellent receiver out-of-band rejection. The transmitter includes provisions for
both on-off keyed (OOK) and amplitude-shift keyed (ASK) modulation. The transmitter employs
SAW filtering to suppress output harmonics.
Absolute Maximum Ratings
Rating
Power Supply and All Input/Output Pins
Non-Operating Case Temperature
Soldering Temperature (10 seconds)
Value
-0.3 to +4.0
-50 to +100
250
Units
V
o
o
403.50 MHz
Hybrid
Transceiver
C
C
Electrical Characteristics (typical values given for 3.0 Vdc power supply, 25
o
C)
Characteristic
Operating Frequency Range
Frequency Drift, -10 to 45 C, 2.2 to 3.7 Vdc
Modulation Type
OOK Data Rate
ASK Data Rate
Receiver Performance, High Sensitivity Mode
Sensitivity, 2.4 kbps, 10-3 BER, AM Test Method
Sensitivity, 2.4 kbps, 10-3 BER, Pulse Test Method
Current, 2.4 kbps (R
PR
= 330 K)
Sensitivity, 19.2 kbps, 10-3 BER, AM Test Method
Sensitivity, 19.2 kbps, 10-3 BER, Pulse Test Method
Current, 19.2 kbps (R
PR
= 330 K)
Sensitivity, 115.2 kbps, 10-3 BER, AM Test Method
Sensitivity, 115.2 kbps, 10-3 BER, Pulse Test Method
Current, 115.2 kbps
Receiver Performance, Low Current Mode
Sensitivity, 2.4 kbps, 10-3 BER, AM Test Method
Sensitivity, 2.4 kbps, 10-3 BER, Pulse Test Method
Current, 2.4 kbps (R
PR
= 1100 K)
Receiver Out-of-Band Rejection, ±5% f
O
R
±5%
1
1
2
3
-104
-98
1.8
80
dBm
dBm
mA
dB
1
1
2
1
1
2
1
1
-109
-103
3.0
-105
-99
3.1
-101
-95
3.8
dBm
dBm
mA
dBm
dBm
mA
dBm
dBm
mA
OOK/ASK
30
115.2
kbps
kbps
o
Sym
f
O
Notes
Minimum
403.35
Typical
Maximum
403.65
±100
Units
MHz
ppm
1