欢迎访问ic37.com |
会员登录 免费注册
发布采购

SCC30D 参数 Datasheet PDF下载

SCC30D图片预览
型号: SCC30D
PDF下载: 下载PDF文件 查看货源
内容描述: 压力传感器 [Pressure sensors]
分类和应用: 传感器压力传感器
文件页数/大小: 5 页 / 104 K
品牌: SENSORTECHNICS [ SENSORTECHNICS GMBH ]
 浏览型号SCC30D的Datasheet PDF文件第1页浏览型号SCC30D的Datasheet PDF文件第3页浏览型号SCC30D的Datasheet PDF文件第4页浏览型号SCC30D的Datasheet PDF文件第5页  
SCC Series
Pressure sensors
PRESSURE SENSOR CHARACTERISTICS
Maximum ratings
(for all devices)
Supply current, I
S
Temperature ranges
Compensated
Operating
Storage
1.5 mA
0 to +50°C
-40 to +85°C
-55 to +125°C
Humidity
Lead temperature (soldering 4 sec.)
Common-mode pressure
0 to 100 %RH
250°C
150 psi
PERFORMANCE CHARACTERISTICS
(individual models) I
S
= 1.0 mA, T
A
= 25°C
1
Operating
pressure
range
0 - 5 psid(g)
0 - 15 psia
0 - 15 psid(g)
0 - 30 psia
0 - 30 psid(g)
0 - 100 psia
0 - 100 psid(g)
Part number
SCC05(D,G)
SCC15A
SCC15(D,G)
SCC30A
SCC30(D,G)
SCC100A
SCC100(D,G)
Proof
pressure
9
20 psi
30 psia
30 psi
60 psia
60 psi
150 psia
150 psi
Accuracy
2
0.50
0.50
0.50
0.50
0.50
0.50
0.50
%
%
%
%
%
%
%
Effect
on span
3
(0 - 50°C)
1.50
1.50
1.50
1.50
1.50
1.50
1.50
%
%
%
%
%
%
%
Effect
on offset
4
(0 - 50°C)
2.00
2.00
2.00
2.00
2.00
2.00
2.00
%
%
%
%
%
%
%
Full-scale
span
1,5
(mV)
25 - 65
30 - 95
40 - 95
60 - 150
60 - 150
85 - 225
85 - 225
PERFORMANCE CHARACTERISTICS
(all models) I
S
= 1.0 mA, T
A
= 25°C
Characteristics
Zero pressure offset
Combined linearity, hysteresis, repeatability
2
Temperature effect on span
3,8
Temperature effect on offset
4,8
Long term stability of offset and span
6
Response time (10 to 90 %)
7
Input impedance
Output impedance
Min.
-30.0
---
---
---
---
---
4.00
4.00
Typ.
-10
0.25
0.25
0.50
0.10
0.10
5.00
5.00
Max.
20.0
0.50
1.50
2.00
---
---
6.50
6.50
Unit
mV
%FSO
%FSO
%FSO
mV
ms
kΩ
kΩ
Specification notes:
1.
2.
Reference conditions: supply current = 1.0 mA, T
A
=25°C, common-mode line pressure = 0 psig, pressure applied to P1, unless
otherwise noted.
Accuracy is the sum of hysteresis and linearity. Hysteresis is the maximum output difference at any point within the operating
pressure range for increasing and decreasing pressure. Linearity refers to the best straight line fit as measured for the
offset, full-scale and 1/2 full-scale pressure at 25°C.
This is the maximum temperature shift for span when measured between 0 and 50°C relative to the 25°C reading.
Typical temperature coefficients for span and resistance are -2200 and +2200 ppm/°C respectively.
This is the maximum temperature shift for offset when measured between 0 and 50°C relative to the 25°C reading.
Span is the algebraic difference between the output voltage at full-scale pressure and the output at zero pressure.
Maximum difference in output at any pressure with the operating pressure range and temperature within 0 to 50°C after:
a) 100 temperature cycles, 0 to 50°C.
b) 1.0 million pressure cycles, 0 psi to full-scale span.
Response time for a 0 psi to full-scale span pressure step change. 10 to 90 % rise time.
Temp. effect on span and offset are guaranteed by design. Therefore these parameters are not 100 % tested.
If the proof pressure is exceeded, even momentarily, the package may leak or burst, or the pressure sensing die may fracture.
Note: The proof pressure for the forward gage of all devices in the D4-package is the specified value or 100 psi whatever is
less. The SCC100D (button package) can be used in forward gage mode only when the pressure doesn't exceed 30 psi,
or no special care ist taken.
3.
4.
5.
6.
7.
8.
9.
2/5
July 2005 / 040