欢迎访问ic37.com |
会员登录 免费注册
发布采购

G2T60 参数 Datasheet PDF下载

G2T60图片预览
型号: G2T60
PDF下载: 下载PDF文件 查看货源
内容描述: 该G2T60系列是在一个非常小的足迹60 W通用输入AC / DC电源。 [The G2T60 Series is a 60 W universal input AC/DC power supply in a very small footprint.]
分类和应用:
文件页数/大小: 2 页 / 411 K
品牌: SLPOWER [ SL Power Electronics ]
 浏览型号G2T60的Datasheet PDF文件第2页  
G2T60 MEDICAL SERIES
Low Power Medical Power Supplies 60 Watt AC/DC Universal Input Power Supply
SINGLE OUTPUT POWER SUPPLY
SUMMARY
• 100 - 240 Vac nominal input range
• Overvoltage and short circuit protection
• Approved to UL, CSA and EN Standards
• Class I input available (GNT60)
The G2T60 Series is a 60 W universal input AC/DC power
supply in a very small footprint. With the medical approvals
the G2T60 is ideal for a variety of medical device applications
including, small single board computers, battery charging, and
running small motors, pumps, and solenoids. The series with
full approval to EN60601-1 and EN60950-1 Standard, improves
design-in time and reduces end system compliance costs.
All specifications are typical at nominal input, full load at 25
O
C unless otherwise stated
SPECIFICATIONS
OUPUT SPECIFICATIONS
Output Power
Total Regulation
Rise time
Transient response
Temperature Coef-
ficient
Overvoltage protection
Short Circuit protection
Power cycling
At turn-on
Main output 50%
0.2 A/ms
Natural Convection
60 Watts max.(65 W with airflow)
See table
2.0 s max.
5% max. dev. 1 ms recovery to
1%
+/-0.03%/oC
125% +/-10%
Yes
INPUT SPECIFICATIONS
Input Voltage range
Input frequency
Input surge current
Ground Leakage current
Input current
Input fuse
Thermistor limited
132 Vac 60 Hz
120 Vac
230 Vac
F1,F2
Universal input
100-240 nom (85 - 264 max)
Vac
55 Hz +/- 10 Hz
37 A max.
100
mA
Input to grounded-
output
1.2 A
0.7 A
3.15 A
EMC INFORMATION
RF emissions
Line freq. harmonics
Voltage fluctuations
ESD
Radiated immunity
Fast Transients (EFT)
Line surge immunity
Conducted immunity
Power freq. mag. field
Voltage dip immunity
EN55011
EN61000-3-2
EN61000-3-3
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
Conducted Class B
Radiated Class A
Complies
Complies
Contact 6 kV
Air 8 kV
3 V/m
2 kV
DIfferential 1 kV
Common 2 kV
3 Vrms
3 A/m
95% dip, 0.5 cycle 60% dip, 5 cycles
30% dip, 25 cycles
ENVIRONMENTAL SPECIFICATIONS
Thermal Performance
operating ambient
(see ratings chart)
non-operating
0 - 50
o
C
Convection cooled
50 -70
o
C ambient
Convection cooled
Relative Humidity
Maximum Altitude
Vibration
Shock
non-condensing
operating / non-operating
5 Hz- 500 Hz
per MIL-STD-810E
0-70
o
C
-40 to +85
o
C
60 Watts
derate to 50%
5% - 95% RH
10,000 ft. / 40,000
ft. max.
2.5 g rms
516.4 part IV
Note: Voltage dip immunity at 100 V is done with a 20% load