型号: GNT60-18
PDF下载: 下载PDF文件 查看货源
内容描述: GNT60 -X- YYYG ,其中X = 12,15, 18,24或48表示的输出电压额定值, -YYY是不相关的安全和G增加值选项表示ROHS符合 [GNT60-X-YYYG, where X = 12, 15, 18, 24 or 48 which represents the output voltage rating, -YYY is value added options not related to Safety and G represents ROHS Compliance]
分类和应用:
文件页数/大小: 2 页 / 122 K
品牌: SLPOWER [ SL Power Electronics ]
 浏览型号GNT60-18的Datasheet PDF文件第2页  
SL POWER ELECTRONICS CORP.
6050 King Drive, Bldg. A
Ventura, CA 93003 + 805-486-4565
Internet: www.slpower.com
GNT60 SERIES
INSTALLATION INSTRUCTIONS
MODEL NUMBERS:
GNT60-X-YYYG, where X = 12, 15, 18, 24 or 48 which represents the output
voltage rating, -YYY is value added options not related to Safety and G represents ROHS Compliance
RATINGS:
Input:
Output:
100-240 V ac, 1.5 A, 50/60 Hz
MODEL
GNT60-12
GNT60-15
GNT60-18
GNT60-24
GNT60-48
40 °C Ambient
Convection Cooled
12 V dc 4.2 A
15 V dc 4.0 A
18 V dc 3.35 A
24 V dc 2.1 A
48 V dc 1.35 A
50 °C Ambient
Convection Cooled
12 V dc 3.35 A
15 V dc 3.7 A
18 V dc 3.1 A
24 V dc 1.9 A
48 V dc 1.25 A
50 °C Ambient
100 LFM Airflow
12 V dc 5.0 A
15 V dc 4.35 A
18 V dc 3.6 A
24 V dc 2.7 A
48 V dc 1.35 A
Notes:
1. Maximum Operating Relative Humidity 96 %, no condensation.
2. Storage: -40 to +85 °C. Units should be allowed to warm-up under non-condensing conditions before application of
power.
CERTIFICATION:
All models are Certified to be in compliance with the applicable requirements of
CSA 22.2 No. 601.1 and EN/IEC/UL 60601-1. In addition, all models (except GNT60-18) are Certified
to be in compliance with EN/IEC/UL/CSA 60950-1.
CLASSIFICATION:
(5.1) Protection against electric shock = Class I
(In accordance with sub-
(5.2) Degree of protection against electric shock = Signal output or intermediate
clause 5 of IEC 60601-1)
(5.3) Protection against harmful ingress of water = Ordinary (no protection)
(5.5) Have not been evaluated for use in the presence of a flammable anaesthetic
mixture with air, oxygen, or nitrous oxide. This evaluation is to be made
on the end equipment by the OEM.
(5.6) Mode of operation = Continuous
SAFETY DECLARATION:
SL Power Electronics Corp. (SLPE) declares under our sole
responsibility that all models listed above are in conformity with the applicable
requirements of EN 60950-1 following the provisions of the Low Voltage Directive
2006/95/EC.
GROUNDING:
The ground connection (quick connect tab) must be bonded to protective earth in the
end equipment. Using this terminal for the system ground is not recommended. A separate dedicated
grounding point should be used.
OUTPUT:
The output is intended for Protectively Earthed Signal Output and Intermediate Circuits
only. The output is not acceptable for patient connection without additional isolation. The DC output is
SELV under normal and single fault conditions.
OVERVOLTAGE PROTECTION:
The output is monitored for an overvoltage condition. In some
applications where an overvoltage condition could result in a hazard as defined in applicable safety
standards, redundant or additional overvoltage protection may be required. Consult factory for details.
41-35510-0001 Rev. D 08/15/2008
Page 1 of 2