MINT1045
45 W Single Output Series
Medical & ITE – Universal Input Power Supply
•
45 W Convection Cooled
•
Universal Input 80--270 Vac
•
Very Small Size: 2” W x 3” L x 1.0” H
•
88% Efficiency Typical
•
Class I or Class II AC Input Applications
•
CSA/EN/IEC/UL60601-1 3
rd
3 Year Warranty
Edition for Medical,
Type BF Capable, Ultra-low Leakage Current
nd
•
CSA/EN/IEC/UL60950-1 2
Edition for ITE
Specifications
AC Input
Input Current
Inrush Current
Input Fuses
Power Factor
Efficiency
Output Power
All Specifications are typical at nominal input, full load at 25
°
C unless otherwise stated.
100-240Vac +/- 10%, 47-63 Hz 1∅ rated
80-270 Vac, 47-440 Hz operational
0.75A @ 120 Vac, 0.45 A @ 240 Vac
37A max., 264 Vac
F1, F2: 1.6A, 250Vac fuses provided internally
Complies with EN61000-3-2, Class A
88% typical
45 W continuous. See chart for operation above 50°C
50 W for 1 minute, 10% duty cycle
Turn On Time
Hold-up Time
Overload Protection
Short Circuit Protection
Overtemperature Protection
500 ms max at 120 Vac
25 ms minimum at 45 W, nominal input (120Vac)
120-150%, auto-reset
Hiccup Mode, 58W typical
Hiccup Mode, 58W typical
Self-Recovering
Overvoltage Protection
Peak Output Power
Ripple and Noise
1% pk-pk max., 20 MHz BW, differential mode
Measured with noise probe directly across output terminals
Output Voltage
See chart
Voltage Adjustability
Not adjustable
Factory set with fixed resistors for higher reliability
Minimum Load
Not required
Total Regulation
Transient Response
Switching Frequency
See chart
500µs max. to 1%, 50% load step
0.2A/µs, 3.5% deviation. typical
Fixed 65 kHz, typical
Medical Safety Standards
CSA/EN/IEC/UL 60601-1, 3
rd
Edition
Methods Of Protection: 2 MOPP for Type BF Applications
ITE Safety Standards
CSA/EN/IEC/UL 60950-1, IEC 2
nd
Edition
See page 2 for detailed safety agency standards
Isolation
Input-Output: 4500Vac, Input-Ground: 1900Vac
Output-Ground: 1900Vac
Leakage Current
12V-28V models: 120Vac: 30µA NC/70µA SFC
264Vac: 75µA NC/130µA SFC
48V model: 120Vac: 90µA NC/120µA SFC
264Vac: 190µA NC/220µA SFC
Operating Temperature
-20° to 70°C
See derating chart
Output Power Derating
Derate output power by 2.5%
per
°C
above 50°C, up to 70°C max
Storage Temperature
-40 to +85°C
Relative Humidity
5% to 95%, non-condensing
MTBF
300K hours per Telcordia
Model Number Key
MINT 1 045 A 12 75 K 01
01 = Standard Model, 02 and higher indicates a m odified m odel.
"K" = Input Connector - 3 pin header
"75" = Output Connector - 4 pin header
Output Voltage: "12" = 12V output, "24" = 24V output, etc.
Model Configuration: "A" = Class I Input; "B" = Class II Input (no ground)
Output Pow er (Watts) - "045" = 45W
# of Outputs
Product Fam ily: "M" = Medical, "I" = Internal, "NT" = New Technology
______________________________________________________________________________________________________________________________
1 of 3
SL Power Electronics Corp
•
6050 King Drive Ventura, CA 93003
•
Phone 805.486.4565
•
Fax 858.712.2040
•
Email: info@slpower.com
•
www.slpower.com