MW116
MEDICAL 60-90 WATT SERIES
Input
Electrical
Topology
Efficiency
EMI
Voltage
100-240 VAC -10%, +10%
Switching - Fixed frequency 100KHz
Up to 89%
Line Frequency
Current
47 - 63 Hz
2.0A max. at 90 VAC input
Complies with EMC Directives
CE Compliant
Protection
Internal primary current fuse,
Inrush limiting
CE
MTBF
100,000 Calculated Hours Minimum
Power Factor Correction
Complies with EN61000-3-2 and
EN61000-3-3
Environmental
Operating Temperature
Storage Temperature
Relative Humidity
Altitude
0º to 40º C with no derating
-30º to +85º C
Output
Combined Line and Load
Voltage Regulation
1% excluding cord
5% to 95% non-condensing
0-10,000 feet
Ripple
1% V p-p max
Transient Response
Holdup Time
Protection
0.5 msec. for 50% load change typical
16 msec min. over entire input range
Cooling
Convectional - non vented case
Foldback overcurrent protection
Short circuit protection
Mechanical
Case and Dimensions
5.08L x 3.05W x 1.67H (in)
129L x 77W x 40H (mm)
Cord and Connector
5V and 6V: 6 ft, 6 conductor, 18 AWG,
Ault #51 connector. Pinout: Pins 1, 2, 3 =
+V; 4, 5 and 6 = common. 9V, 12V, 13.5V,
15V, 16V: 6 ft, 4 conductor, 18 AWG,
Ault #51 connector. Pinout: Pins 1,4=+V;
3,6=common. All with overmolded
Ferrite core. 18V, 24V, 48V,: 6 ft, 2 con-
ductor, 18 AWG with overmolded Ferrite
core, Ault #3 connector, center positive.
Case Material
Weight
94V0 polycarbonate; black is standard
color; other colors available - extra
charges may apply
13 ounces, 370 grams
(excluding cord)
Safety Approvals
Agency Listings
TUV EN60601-1; cURus 2601;
Australian, Japanese certification
available - extra fees apply
Input Configuration
Connector
IEC320
w/ground
C14
IEC320
w/o ground
C18
(F)
(Q)
In case IEC320 type: F=Class I C14 inlet Q=Class II C18 inlet
Specify the Input Configuration Code in your order.
A U LT I N C O R P O R AT E D
7105 Northland Terrace | Minneapolis, Minnesota 55428 | USATEL (1) 763-592-1900
FAX (1) 763-592-1911 | E-MAIL info@aultinc.com | WEB www.aultinc.com
© Copyright 2005 Ault Inc.