 浏览型号AQ0400MU36EABF的Datasheet PDF文件第5页浏览型号AQ0400MU36EABF的Datasheet PDF文件第6页浏览型号AQ0400MU36EABF的Datasheet PDF文件第7页浏览型号AQ0400MU36EABF的Datasheet PDF文件第8页浏览型号AQ0400MU36EABF的Datasheet PDF文件第9页浏览型号AQ0400MU36EABF的Datasheet PDF文件第10页浏览型号AQ0400MU36EABF的Datasheet PDF文件第11页浏览型号AQ0400MU36EABF的Datasheet PDF文件第13页  
85-264 V  
12/24/36/48V Semi-reg  
400 W Series  
Medical  
AC Input:  
DC Output:  
Power:  
RMS  
.
Grade:  
Technical Specification  
INSTALLATION INSTRUCTIONS  
Output:ꢀꢀRefer to the Connector Details section for output con-  
nector wiring and signal I/O functionality. Refer to nameplate  
label for output current ratings. Main DC output (Vout+, Vout-)  
pins should use 20 AWG (0.5mm2) wire size. Individual main  
output pins should not be loaded to more than 5.5 A. For cur-  
rents greater than 5.5 A, multiple main output pins/wires must be  
used in parallel. All signal I/O pins are referenced to Vout-.  
General:ACuQor AC/DC power supplies are intended for use  
as components in medical and industrial equipment. ACuQor  
units must be properly installed within end use equipment before  
they can be safely applied as described in this document. The  
suitability of the ACuQor/equipment combination must be verified  
through end product investigation.  
Mounting:ꢀꢀRefer to the Mechanical Drawings section. ACuQor  
units are provided with threaded stainless-steel stand-offs or  
inserts for mounting. This mounting hardware is internally con-  
nected to the input connector protective-earth terminal for func-  
tional-earth EMC control. Any orientation (vertical, horizontal,  
etc.) may be used. Adequate air space should be provided over  
the fan intake (top) and exhaust (sides) to allow for exchange of  
cooling air. ACuQor is designed for a pollution degree 2 environ-  
ment. The suitability of the enclosed ACuQor mechanical assem-  
blies must be verified through end product investigation.  
EMC:ꢀꢀACuQor products have been tested to the EMC specifica-  
tions listed in the Electrical Characteristics section. However, end  
use equipment must be tested to verify EMC compliance.  
.
Hipot Testing: ACuQor products are rated for Hipot testing  
levels of 1500 Vac input to protective-earth, 1500 Vac output to  
protective-earth, and 4000 Vac input to output. When performing  
the 4000 Vac input to output test, the test voltage must be bal-  
anced evenly 2000 Vac input and output to protective-earth. Two  
oppositely phased test voltage sources or a single test voltage  
source with external balancing impedances (capacitors) may be  
used to prevent overstressing input or output to protective-earth  
insulation per IEC60601-1 2005 sub clause 8.8.1 and IEC60601-1  
1990 sub clause 20.4g.  
Encased models:A minimum of 5 mm electrical clearance  
should be allowed from the connector ends of encased models.  
Input:Refer to the Connector Details section for input connector  
wiring. ACuQor products require a single phase AC power source  
of 100-240V 50/60Hz nominal. Refer to nameplate label for input  
current ratings. A protective-earth connection is also required.  
Minimum wire size of 18 AWG (0.8mm2) is recommended. Both  
sides of the AC line are internally fused (see table for specific  
Patient Contact: ACuQor models include versions designed  
for BF and CF patient contact application per IEC60601-1. These  
ACuQor models provide reinforced insulation at the DC output  
voltage level and basic insulation at the 240 Vac level from output  
to protective-earth. Note that equipment and wiring may add to  
system leakage currents so that the end product must be tested  
for compliance. Refer to the Electrical Characteristics section for  
typical ACuQor input and output leakage currents. In addition,  
ACuQor defibrillation rated models comply with the minimum  
output to protective-earth creepage/clearance requirement and  
defibrillator pulse test of IEC60601-1.  
models). These fuses are not user replaceable  
.
MODEL  
AQ0300  
AQ0400  
AQ0500  
Input Fuses (in Both AC Lines)  
Littelfuseꢀ6.3Aꢀ250Vꢀ21606.3XEP  
Littelfuseꢀ6.3Aꢀ250Vꢀ21606.3XEP  
Littelfuseꢀ10.0Aꢀ250Vꢀ216010XEP  
Fuses Total  
2
2
2
Product # AQ0400MUxx Series  
Phone 1-888-567-9596  
www.synqor.com  
Doc.# 005-005045 Rev. A  
09/08/08  