欢迎访问ic37.com |
会员登录 免费注册
发布采购

MPS-65-24 参数 Datasheet PDF下载

MPS-65-24图片预览
型号: MPS-65-24
PDF下载: 下载PDF文件 查看货源
内容描述: 65W单输出医用型 [65W Single Output Medical Type]
分类和应用:
文件页数/大小: 2 页 / 344 K
品牌: TOTAL-POWER [ TOTAL POWER INTERNATIONAL ]
 浏览型号MPS-65-24的Datasheet PDF文件第2页  
TOTAL POWER INT'L
65W Single Output Medical Type
Features :
Low leakage current
0.3mA
MPS-65
series
Universal AC input / Full range
Protections:Short circuit/Over load/Over voltage
Cooling by free air convection
100% full load burn-in test
Fixed switching frequency at 45KHz
3 years warranty
SPECIFICATION
MODEL
DC VOLTAGE
RATED CURRENT
CURRENT RANGE
RATED POWER
OUTPUT
OUTPUT POWER (max.)
VOLTAGE ADJ. RANGE
MPS-65-3.3
3.3V
12A
0 ~ 15.2A
39.6W
MPS-65-5
5V
12A
0 ~ 13.8A
60W
100mVp-p
MPS-65-7.5
7.5V
8A
0 ~ 9.6A
60W
100mVp-p
MPS-65-12
12V
5.2A
0 ~ 6A
62.4W
100mVp-p
MPS-65-13.5 MPS-65-15
13.5V
4.7A
0 ~ 5.4A
63.5W
100mVp-p
15V
4.2A
0 ~ 4.8A
63W
100mVp-p
MPS-65-24
24V
2.7A
0 ~ 3A
64.8W
100mVp-p
MPS-65-27
27V
2.4A
0 ~ 2.7A
64.8W
100mVp-p
MPS-65-48
48V
1.35A
0 ~ 1.5A
64.8W
100mVp-p
72W(+3.3V:50W;+5V:69W)with 18CFM min. Forced air convection
6.75 ~ 8.25V 10.8 ~ 13.2V 12.2 ~ 14.85V 13.5 ~ 16.5V 21.6 ~ 26.4V 24.3 ~ 29.7V 43.2 ~ 52.8V
3.0%
2.0%
2.0%
2.0%
2.0%
2.0%
2.0%
RIPPLE & NOISE (max.)
Note.2
80mVp-p
2.97 ~ 3.63V 4.5 ~ 5.5V
3.0%
VOLTAGE TOLERANCE
Note.3
3.0%
LINE REGULATION
1.0%
1.0%
LOAD REGULATION
3.0%
3.0%
1.0%
3.0%
1.0%
2.0%
1.0%
2.0%
1.0%
2.0%
1.0%
2.0%
1.0%
2.0%
1.0%
2.0%
SETUP, RISE, HOLD TIME
VOLTAGE RANGE
FREQUENCY RANGE
INPUT
EFFICIENCY(Typ.)
AC CURRENT
INRUSH CURRENT(max.)
LEAKAGE CURRENT
OVER LOAD
PROTECTION
800ms, 30ms, 20ms/230VAC
800ms, 30ms, 12ms/115VAC at full load
90 ~ 264VAC
47 ~ 440Hz
66%
1.6A/115VAC
127 ~ 370VDC
74%
76%
0.9A/230VAC
40A/230VAC
77%
78%
79%
80%
80%
80%
COLD START 20A/115VAC
<0.3mA / 264VAC
73 ~ 105W (3.3V:51 ~ 75W)(5V:70 ~ 105W)
rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
3.8 ~ 4.46V
5.75 ~ 6.75V 8.63 ~ 10.1V 13.8 ~ 16.2V 15.5 ~ 18.2V 17.25 ~ 20.25V 27.6 ~ 32.4V 31 ~ 36.45V
55.2 ~ 64.8V
OVER VOLTAGE
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT
STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
WITHSTAND VOLTAGE
Protection type : Hiccup mode, recovers automatically after fault condition is removed
-10 ~ +60 (Refer to output load derating curve)
20 ~ 90% RH non-condensing
-20 ~ +85
, 10 ~ 95% RH
0.04%/ (0 ~ 50
10 ~ 500Hz, 2G 10min./1cycle, Period for 60min.each along X, Y, Z axes
UL2601-1, TUV EN60601-1, IEC601-1 Approved
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:0.5KVAC
SAFETY & ISOLATION RESISTANCE
I/P-O/P, I/P-FG:100M Ohms/500VDC
EMC
EMI CONDUCTION & RADIATION
Compliance to EN55011 (CISPR11) Class B
(Note 4)
HARMONIC CURRENT
EMS IMMUNITY
MTBF
Compliance to EN61000-3-2,-3
Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN60601-1-2 Medical level, criteria A
OTHERS
DIMENSION
PACKING
359.7Khrs min. MIL-HDBK-217F (25
127*76.2*42mm (L*W*H)
0.23Kg; 54pcs/14.6Kg/1.35CUFT
)
NOTE
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives.
5. Mounting holes M1 and M2 should be grounded for EMI purposes.