欢迎访问ic37.com |
会员登录 免费注册
发布采购

PM40AL 参数 Datasheet PDF下载

PM40AL图片预览
型号: PM40AL
PDF下载: 下载PDF文件 查看货源
内容描述: 40瓦医疗电源 [40 WATT MEDICAL POWER SUPPLIES]
分类和应用: 医疗
文件页数/大小: 2 页 / 207 K
品牌: TRUMPOWER [ Tumbler Technologies + TRUMPower ]
 浏览型号PM40AL的Datasheet PDF文件第2页  
40 WATT MEDICAL POWER SUPPLIES
DESCRIPTION
The PM40AL series of compact, open PCB constructed, AC-DC
switching power supplies are capable of delivering 40 watts of
continuous output power. They operate at 85 to 264 VAC
input voltage without the need of voltage selection. They are
ideally suited for use in medical equipment, safety systems and
monitoring equipment, not for patient contact.
PM40AL SERIES
(MDD)
F E ATU R E S
Low safety ground leakage current
100% burn-in
W i d e i n p u t r a n g e 8 5 t o 2 6 4 VA C
Input surge current protection
Overvoltage protection
Overcurrent protection
Open PCB construction
I N P U T S P E C I F I C ATI O N S
Input voltage:
Input frequency:
Input current:
Leakage current:
85 to 264 VAC
47 to 63 Hz
S a f e t y St a n d a r d A p p ro v a ls
UL 2601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
1.20A(rms)for 115VAC
0.70A(rms)for 230VAC
150uA max. @ 230VAC, 50Hz
90 uA max. @ 115VAC, 60Hz
GENERAL SPECIFICATIONS
OUTPUT SPECIFICATIONS
Output voltage/current:
Total output power:
Ripple and noise:
Overvoltage protection:
See rating chart
40 watts maximum
1﹪ peak to peak maximum
Provided on output #1 only; set at
112 - 132﹪ of its nominal output
voltage
All outputs protected to short
circuit conditions
All outputs ±0.04﹪ /℃ maximum
Maximum excursion of 4﹪ or
better on all models, recovering to
1﹪ of final value within 500us
after a 25﹪ step load change
Switching frequency:
Efficiency:
36KHz ±5KHz
70﹪ minimum on single output models
with Vo≧ 15V, 65% minimum on the
others
10 msec minimum at 110VAC
±0.5﹪ maximum at full load
Hold-up time:
Line regulation:
Inrush current:
Overcurrent protection:
Temperature coefficient:
Transient response:
E N V I R O N M E N TA L S P E C I F I C ATI O N S
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
0℃ to +70℃
-40℃ to +85℃
5﹪to 95﹪non-condensing
Derate from 100﹪ at +50℃ linearly to
50% at +70℃
15 amps @ 115VAC, or 30 amps @
230VAC, at 25℃ cold start
Withstand voltage:
4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
MTBF:
600,000 hours minimum at full load at
25℃ ambient, calculated per
MIL-HDBK-217F
EMC Performance (EN60601-1-2: 2001)
Class B conducted, class B radiated
EN55011
Class B conducted, class B radiated
FCC
Class B conducted, class B radiated
VCCI
Harmonic distortion, class A and D
EN61000-3-2
Line flicker
EN61000-3-3
ESD, ± 8KV air and ± 6KV contact
EN61000-4-2
Radiated immunity, 3V/m for 80~2500MHz
EN61000-4-3
Fast transient/burst, ± 2KV
EN61000-4-4
Surge, ± 1KV diff., ± 2KV com.
EN61000-4-5
Conducted immunity, 3Vrms
EN61000-4-6
Magnetic field immunity, 3A/m
EN61000-4-8
Voltage dips, 30% reduction for 500ms,
EN61000-4-11
60% reduction for 100ms and >95%
reduction for 10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616
sales@trumpower.com