TRUMPower
D E SCR IPTION
The TM300 series comprising single and multiple
output models for 200 to 300 watts of continuous
output power is specially designed for medical
applications. They operate at 90 to 264VAC input
voltage without the need of a selector strap. All
auxiliary outputs are with magnetic amplifier to
keep regulation. The units are constructed on a
printed circuit board with a U-bracket for
mechanical
support
and
heat
sinking.
A
cover-and-fan assembly can be added during
manufacturing.
300 WATT MEDICAL POWER SUPPLIES
TM300 SERIES
( LV D )
RoHS
F EATU R E S
EN61000-3-2 class A and D compliant
Pow er Factor 0.98 typical
O v e r v o l ta g e p r o t e c t i o n
Short-circuit protection
Pow er Fail Detect (PFD)
100% burn-in at full rated load
Remote sense on output #1 and output #2
Remote inhibit – TTL high disables output
C o m p l i a n t w i t h R o H S r e q u i r e m e n ts
Saf et y Sta n da rd A ppr o va ls:
UL 60601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
INPUT SPECIFICATIONS
Input voltage :
Input frequency:
Input current :
90 to 264VAC
47 to 63Hz
4.7A (rms) for 115VAC
2.3A (rms) for 230VAC
Earth leakage current:
100µA max. @115VAC, 60Hz
220µA max. @ 230VAC, 50Hz
E N V IR O N M ENTA L SP E C IF ICATI O N S
Operating temperature
Storage temperature
Relative humidity
Derating
Cooling
0ºC to +70ºC
-40ºC to +85ºC
5% to 95% non-condensing
Derate from 100% at +50ºC
linearly to 50% at +70ºC
200/250/300 watts continuous
output power at 35 CFM forced
air cooling or 100/125/150 watts
at convention cooling
OUTPUT SPECIFICATIONS
Output voltage/current:
Total output power:
Ripple and Noise:
Overvoltage protection:
See rating chart
See rating chart
2% peak to peak maximum
Provided on output #1 only; set
at 115–140% of its nominal
output voltage
All outputs protected
to short circuit conditions
All outputs ±0.04%/ºC
maximum
Maximum excursion of 4%
or better on all models,
recovering to 1% of final value
within 500µs after a 25% step
load change
TTL logic high for normal
operation and TTL logic low
upon loss of input power.
This signal appears at least
1ms prior to master output
dropping 5% below its nominal
value. This signal also
provides a minimum delay of
100ms after master output is
within regulation.
Requires an external TTL
high level signal to inhibit
outputs for standard models.
Switching frequency:
Power factor:
Efficiency:
Hold-up time:
Line regulation:
Inrush current:
GENERAL SPECIFICATIONS
KHz ±10KHz
0.98 typical
70% minimum on all models
12 msec minimum at 110VAC
±0.2% maximum at full load
30 amps @ 115VAC or 60 amps
@230VAC, at 25ºC cold start
4000VAC from input to output
Withstand voltage:
1500VAC from input to ground
500VAC from output to ground
300,000 hours minimum at full load
MTBF:
at 25ºC ambient, calculated per
MIL-HDBK- 217F
EMC Performance (EN60601-1-2: 2001)
EN55011:
Class B conducted, Class B
radiated
EN61000-3-2:
Harmonic distortion, Class A and D
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
Radiated immunity, 3V/m
EN61000-4-4:
Fast transient/burst, ±2KV
EN61000-4-5:
Surge, ±1KV diff, ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
EN61000-4-11:
Voltage dips, 30% reduction for
500ms, 60% reduction for 100ms
and >95% reduction for 10ms
Overcurrent protection:
Temperature coefficient:
Transient response:
PFD signal:
Remote inhibit:
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616
Fax: 408-988-6622
email: sales@trumpower.com
Website: www.TRUMPower.com