D E SCR IPTION
30-40 WATT MEDICAL POWER SUPPLIES
TM41 SERIES
( LV D )
The TM41 series of compact, open PCB constructed,
AC-DC switching power supplies are capable of
delivering 30 to 40 watts of continuous output power.
They operate at 85 to 264VAC input voltage without
the need of voltage selection. They are ideally suited
for use in medical equipments, safety systems and
monitoring equipments.
RoHS
F EATU R E S
Low safety ground leakage current
100% burn-in
8 5 t o 2 6 4 VA C w i d e i n p u t r a n g e
Input surge current protection
O v e r v o l ta g e p r o t e c t i o n
Overcurrent protection
Open PCB construction
C o m p l i a n t w i t h R o H S r e q u i r e m e n ts
Saf et y Sta n da rd A ppr o va ls:
UL 2601-1
CSA C22.2 No. 601.1
IN PUT S PE C IF ICAT ION S
I n p u t v o l ta g e
Input frequency
Input current
Earth leakage current
8 5 t o 2 6 4 VA C
11 0 t o 2 8 0 V D C
47 to 63Hz
1 . 1 A
r m s
f o r 11 5 VA C
0 . 6 A
r m s
f o r 2 3 0 VA C
9 0 µ A m a x . @ 11 5 VA C , 6 0 H z
1 5 0 µ A m a x . @ 2 3 0 VA C , 5 0 H z
TÜV EN60601-1
GENERAL SPECIFICATIONS
Switching frequency:
42KHz ±5KHz
Efficiency:
70% minimum on single output
model with Vo
12V, 68%
minimum on the others
20 msec minimum at 110VAC
±0.5% maximum at full load
12 amps @ 115VAC or 24 amps
@ 230VAC, at 25ºC cold start
4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
OUTPUT SPECIFICATIONS
Output voltage/current:
Total output power:
Ripple and noise:
Over voltage protection:
See rating chart
See rating chart
1% peak to peak max.
Provided on output; set at
112 – 132% of its nominal
output voltage
The output protected to short
circuit conditions
All outputs ±0.04%/ºC
i
Maximum excursion of 4% or
better on all models,
recovering to 1% of final value
within 500µs after a 25% step
load change
Hold-up time:
Line regulation:
Inrush current:
Withstand voltage:
Over current protection:
Temperature coefficient:
Transient response:
600,000 hours minimum at full load
at 25ºC ambient, calculated per
MIL-HDBK- 217F
EMC Performance (EN60601-1-2: 2001)
E N 5 5 0 11 :
FCC:
VCCI:
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
EN61000-4-3:
EN61000-4-4:
EN61000-4-5:
EN61000-4-6:
EN61000-4-8:
E N 6 1 0 0 0 - 4 - 11 :
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Harmonic distortion, Class A and D
Line flicker
ESD, ±8KV air and ±6KV contact
Radiated immunity, 3V/m for 80-2500MHz
Fast transient/burst, ±2KV
Surge, ±1KV diff, ±2KV com.
Conducted immunity, 3Vrms
Magnetic field immunity, 3A/m
Voltage dips, 30% reduction for 500ms,
60% reduction for 100ms and >95%
reduction for 10ms
MTBF:
E N V IR O N M ENTA L SP E C IF ICATI O N S
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
0ºC to +70ºC
-40ºC to +85ºC
5% to 95% non-condensing
Derate from 100% at +50ºC
linearly to 50% at +70ºC
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616
Fax: 408-988-6622
email: sales@trumpower.com