TRUMPower
D E SCR IPTION
The TM650 series comprising single and
multiple output models for 650-700 watts of
continuous output power is specially designed
for medical applications not for life support.
They operate at 90 to 264VAC input voltage
without the need of a selector strap. The units
are constructed on a printed circuit board with a
U-bracket for mechanical support and heat
sinking. A cover-and-fan assembly can be
added during manufacturing.
650 WATT MEDICAL POWER SUPPLIES
TM650 SERIES
(LVD)
RoHS
F EATU R E S
EN61000-3-2 class A and D compliant
Pow er Factor 0.98 typical
O v e r v o l ta g e p r o t e c t i o n
Short-circuit protection
Thermal protection
Pow er Fail Detect (PFD)
100% burn-in at full rated load
Remote sense on output #1 and output #2
Remote inhibit – TTL high disables output
C o m p l i a n t w i t h R o H S r e q u i r e m e n ts
Saf et y Sta n da rd A ppr o va ls:
UL 60601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
Input current:
Earth leakage current:
90 to 264VAC
47 to 63Hz
10A(rms)for 115VAC
5A(rms)for 230VAC
100µA max. @ 115VAC, 60Hz
165µA max. @ 230VAC, 50H z
E N V IR O N M ENTA L SP E C IF ICATI O N S
Operating temperature
0ºC to +70ºC
Storage temperature
-40ºC to +85ºC
Relative humidity
5% to 95% non-condensing
Derating
Derate from 100% at +50ºC
linearly to 50% at +70ºC
OUTPUT SPECIFICATIONS
Output voltage/current:
Total output power:
Ripple and noise:
See rating chart
See rating chart
2% peak to peak maximum
GENERAL SPECIFICATIONS
Switching frequency:
Power factor:
Efficiency:
Hold-up time:
Line regulation:
Inrush current:
Withstand voltage:
70KHz ±10KHz
0.98 typical
80% minimum on all models
12 msec minimum at 110VAC
±0.2% maximum at full load
50amps @115VAC or 100 amps
@ 230VAC at 25ºC cold start
Overvoltage protection:
Provided on output #1 only; set
at 115 –140% of its nominal output
voltage
Overcurrent protection:
All outputs protected to short
circuit conditions
Temperature coefficient:
All outputs ±0.04%/ºC maximum
Transient response:
Maximum excursion of 4% or
better on all models, recovering to
1% of final value within 500µs
after a 25% step load change
TTL logic high for normal
operation and TTL logic low upon
loss of input power. This signal
appears at least 1ms prior to
master output dropping 5% below
its nominal value. This signal also
provides a minimum delay of 100
ms after master output is within
regulation.
Requires an external TTL high
level signal to inhibit outputs for
standard models.
PFD signal:
Remote inhibit:
4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
300,000 hours minimum at full load
MTBF:
at 25ºC ambient, calculated per
MIL-HDBK- 217F
EMC Performance (EN60601-1-2)
EN55011:
Class B conducted, Class A radiated
EN61000-3-2:
Harmonic distortion, Class A and D
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
Radiated immunity, 3V/m
EN61000-4-4:
Fast transient/burst, ±2KV
EN61000-4-5:
Surge, ±1KV diff, ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
EN61000-4-11:
Voltage dips, 30% reduction for
500ms, 60% reduction for 100ms
and >95% reduction for 10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616
Fax: 408-988-6622
email: sales@trumpower.com
Website: www.TRUMPower.com