MPT-120A 参数 Datasheet PDF下载

MPT-120A图片预览
型号: MPT-120A
PDF下载: 下载PDF文件 查看货源
内容描述: 1 〜 4输出医用型 [1~4 Output Medical Type]
分类和应用:
文件页数/大小: 3 页 / 291 K
品牌: XPPOWER [ XP POWER LIMITED ]
 浏览型号MPT-120A的Datasheet PDF文件第2页浏览型号MPT-120A的Datasheet PDF文件第3页  
1~4 Output Medical Type
120W Medical
Features
Universal AC input / Full range
Low leakage current <180 A
Protections:Short circuit/Over load/Over voltage
UL60601-1 medical safety approved
With power good and fail signal output (Optional)
100% full load burn-in test
Fixed switching frequency at 45KHz
3 years warranty
series
IEC60601-1
Specification
AC INPUT VOLTAGE
90~264 VAC, 47~440Hz / 127~370VDC.
AC INPUT CURRENT
(Typ.)
Maximum input current 2.9A at 115VAC, 60Hz or
1.7A at 230VAC, 60Hz with 100% output load.
INRUSH CURRENT
(Typ.)
Inrush current is less than 22A at 115VAC or less than
45A at 230VAC under cold start conditions. Limiting
provided by internal thermistors.
SETUP, RISE TIME
MPS-120:
800ms, 20ms / 230VAC at full load
2000ms, 50ms / 115VAC at full load
MPD,T,Q-120:500ms, 20ms / 230VAC at full load
1200ms, 50ms / 115VAC at full load
HOLD-UP TIME
(Typ.)
80ms / 230VAC at full load
14ms / 115VAC at full load
LEAKAGE CURRENT
Leakage current is less than 180 A at 264VAC
DC OUTPUT ADJ. RANGE
DC output voltage (or Ch1 of multiple output models)
can be adjusted between -5%~+10% rated output
voltage by potential meter.
OVER LOAD PROTECTION
Fully protected against short circuit and output overload.
The hiccup type protection will be activated at 110~150%
(For MPD,T,Q-120),120~160%(For MPS-120) rated load
and recovers automatically after fault condition is removed.
OVER VOLTAGE PROTECTION
Provided on output channel 1 only at 115%~135%
rated output voltage. (120%~140% for MPS-120-15/24/48).
Output will be shut down when this protection is activated.
POWER GOOD / FAIL SIGNAL (OPTIONAL)
TTL logic high for power good and TTL low for
power fail. When the output voltage reaches 90% of
rated value, a +5V TTL signal will be sent out with a
10~500ms delay; At least 1ms before the output
voltage goes below 90% of the rated value, the TTL
signal will be turned off.
* MPS-120-3.3 does not have this optional function.
WORKING TEMP.
Whole series can operate from -20~70
to the derating curves.
WORKING HUMIDITY
20~90% RH non-condensing.
STORAGE TEMP., HUMIDITY
-40~+85 , 10~90% RH
. Please refer
TEMP. COEFFICIENT
0.04%/ on all outputs at full load between 0~50
of ambient temperature.
VIBRATION
2G of acceleration, vibrating frequency adjust from 10Hz ~500Hz within a
10-minute cycle. 6 testing cycles (60 minutes) each along X, Y, Z axes.
SAFETY STANDARDS
Medical
:
UL60601-1, TUV EN60601-1 approved
Commercial : Also design refer to UL60950-1, TUV EN60950-1
WITHSTAND VOLTAGE
4000VAC between input and output
1500VAC between input and F.G.
500VAC between output and F.G.
ISOLATION RESISTANCE
>100M Ohms for I/P-O/P, I/P-FG, O/P-FG by using 500VDC test voltage.
EMI COMPLIANCE
EMI Specifications
Conducted & Radiation
Harmonic distortion
Voltage flicker
EMS COMPLIANCE
EMS Specification
ESD air
ESD contact
RF field susceptibility
EFT(Electrical Fast Transient)/Burst
Lightning/Surge
Conducted RF susceptibility
Magnetic field immunity
Voltage dip, interruption
Digital phone carrier immunity
MTBF
262,100 hours min. at full load and 25
calculated per MIL-HDBK-217F.
DIMENSION (L*W*H)
177.8x107.95x35.5mm or 7"x4.25"x1.4"
PACKING
0.55Kg; 24pcs/14.5Kg/1.04CUFT
Compliance Level
EN55011, Class B
EN55022, Class B
EN61000-3-2
EN61000-3-3
Compliance Level
EN61000-4-2, Level 3, 8KV
EN61000-4-2, Level 2, 4KV
EN61000-4-3, Level 2, 3V/m
Level 3, 10V/m
EN61000-4-4, Level 2, 1KV/5KHz
Level 3, 2KV/5KHz
EN61000-4-5, Level 4, 2KV/Line-Line
4KV/Line-Earth
EN61000-4-6, Level 2, 3Vrms/m
Level 3, 10Vrms/m
EN61000-4-8, Level 2, 3A/m
Level 3, 10A/m
EN61000-4-11,Compliance
ENV50204, Level 2, 3V/m, 900MHz
Level 3, 10A/m, 900MHz
of ambient temperature,
File Name:120W MEDICAL-SPEC 2006-04-13