欢迎访问ic37.com |
会员登录 免费注册
发布采购

2002 参数 Datasheet PDF下载

2002图片预览
型号: 2002
PDF下载: 下载PDF文件 查看货源
内容描述: 万用表高性能 [MULTIMETER HIGH PERFORMANCE ]
分类和应用:
文件页数/大小: 2 页 / 187 K
品牌: ETC [ ETC ]
 浏览型号2002的Datasheet PDF文件第1页  
2001
2002
Ordering Information
2001 High Performance 7
1
2
-Digit
DMM with 8K Memory
2001/MEM1
High Performance 7
1
2
-Digit
DMM with 32K Memory
2001/MEM2
High Performance 7
1
2
-Digit
DMM with 128K Memory
2002 High Performance 8
1
2
-Digit
DMM with 8K Memory
2002/MEM1
High Performance 8
1
2
-Digit
DMM with 32K Memory
2002/MEM2
High Performance 8
1
2
-Digit
DMM with 128K Memory
2000-SCAN
10-Channel Scanner Card
2001-SCAN
10-Channel Scanner Card
with two high-speed
channels
2001-TCSCAN
9-Channel Thermocouple
Scanner Card
These products are available
with an Extended Warranty.
Accessories Supplied
Model 8605 High Performance
Modular Test Leads, user’s man-
ual, option slot cover, and full
calibration data.
7
1
2
-Digit High Performance Multimeter
8
1
2
-Digit High Performance Multimeter
2001 Condensed Specifications
DC VOLTS
RANGE
200 mV
2 V
20 V
200 V
1000 V
ACCURACY
±(ppm of reading
INPUT
+ ppm of range)
RESOLUTION RESISTANCE
90 Days
10 nV
>10 GΩ
25 + 6
100 nV
>10 GΩ
18 + 2
1 µV
>10 GΩ
18 + 4
10 µV
10 MΩ ±1%
27 + 3
100 µV
10 MΩ ±1%
31 + 6
ACCURACY
±(ppm of reading
CURRENT + ppm of range)
SOURCE
90 Days
9.2 mA
52 + 7
0.98 mA
36 + 7
0.98 mA
33 + 4
89 µA
32 + 4
7 µA
72 + 4.5
770 nA
110 + 4.5
70 nA
560 + 4.5
4.4 nA
10000 + 100
4.4 nA
20000 + 100
ACCURACY
MAXIMUM ±(ppm of reading
BURDEN
+ ppm of range)
VOLTAGE
90 Days
0.25 V
300 + 25
0.31 V
300 + 20
0.4 V
300 + 20
0.5 V
300 + 20
1.5 V
600 + 20
2002 Condensed Specifications
DC VOLTS
RANGE
200 mV
2 V
20 V
200 V
1000 V
ACCURACY
±(ppm of reading
INPUT
+ ppm of range)
RESOLUTION RESISTANCE
90 Days
1 nV
>100 GΩ
15 + 8
10 nV
>100 GΩ
6 + 0.8
100 nV
>100 GΩ
6 + 0.15
1 µV
10 MΩ ±1%
14 + 2
10 µV
10 MΩ ±1%
14 + 0.4
ACCURACY
±(ppm of reading
CURRENT + ppm of range)
SOURCE
90 Days
7.2 mA
15 + 6
960 µA
15 + 4
960 µA
7 + 0.4
96 µA
7 + 0.4
9.6 µA
29 + 0.8
1.9 µA
53 + 0.5
1.4 µA
175 + 0.6
1.4 µA
500 + 3
1.4 µA
2000 + 15
ACCURACY
MAXIMUM ±(ppm of reading
BURDEN
+ ppm of range)
VOLTAGE
90 Days
0.25 V
275 + 25
0.3 V
275 + 20
0.35 V
275 + 20
0.35 V
300 + 20
1.1 V
600 + 20
Model 2001 and 2002 Specifications
RESISTANCE
RANGE
20
200
2 kΩ
20 kΩ
200 kΩ
2MΩ
20MΩ
200MΩ
1GΩ
RESOLUTION
1 µΩ
10 µΩ
100 µΩ
1mΩ
10mΩ
100mΩ
1
10
100
RESISTANCE
RANGE
20
200
2 kΩ
20 kΩ
200 kΩ
2MΩ
20MΩ
200MΩ
1GΩ
RESOLUTION
100 nΩ
1 µΩ
10 µΩ
100 µΩ
1mΩ
10mΩ
100mΩ
1
10
DC AMPS
RANGE
200 µA
2 mA
20 mA
200 mA
2 A
RESOLUTION
10 pA
100 pA
1 nA
10 nA
100 nA
DC AMPS
RANGE
200 µA
2 mA
20 mA
200 mA
2 A
RESOLUTION
10 pA
100 pA
1 nA
10 nA
100 nA
GENERAL/STANDARDS COMPLIANCE
POWER
Voltage:
90–134V and 180–264V universal self-selecting.
,
Frequency:
50Hz, 60Hz, or 400Hz self-identifying.
Consumption:
<55 VA.
ENVIRONMENT
Operating Temperature:
0° to 50°C.
Storage Temperature:
–40° to +70°C.
Humidity:
80% R.H., 0° to 35°C.
Altitude:
4,500m (15,000 ft) operating; 12,000m (40,000
ft.) non-operating.
PHYSICAL
Case Dimensions:
90mm high
×
214mm wide
×
369mm
deep (3
1
2
in
×
8
1
2
in
×
14
1
2
in).
Unit Weight:
4.2kg (9.2 lbs).
STANDARDS
EMI/RFI:
Conforms to VDE 0871B (per Vfg 1046/1984),
IEC 801-2, FCC part 15 Class B, CISPR-22 (EN55022).
Safety:
Conforms to IEC348, CAN/CSC22.2 No. 231,
MIL-T-28800E1.
FREQUENCY COUNTER
RANGE:
1Hz–15MHz.
ACCURACY:
±(0.03% of reading).
DC IN-CIRCUIT CURRENT
RANGE:
100µA to 12A.
ACCURACY:
±(5% + 2 counts) over 2 years.
TRACE RESISTANCE:
1mΩ to 10Ω typical.
D I G I T A L M U LT I M E T E R S
TEMPERATURE
Built-in linearization for J, K, N, T, E, R, S, B thermocouple types
to ITS-90 and 100Ω platinum RTDs.
For complete specifications, refer to the 2001 or 2002
Technical Data Book.
For more information, request the Model
2001 and 2002 Technical Specifications
books.
1.888.KEITHLEY
(U.S. only)
www.keithley.com
A
G R E A T E R
M E A S U R E
O F
C O N F I D E N C E