欢迎访问ic37.com |
会员登录 免费注册
发布采购

26PC10K0D2 参数 Datasheet PDF下载

26PC10K0D2图片预览
型号: 26PC10K0D2
PDF下载: 下载PDF文件 查看货源
内容描述: 温度。补偿和校准压力传感器 [Temp. compensated and calibrated pressure sensors]
分类和应用: 传感器压力传感器
文件页数/大小: 4 页 / 55 K
品牌: SENSORTECHNICS [ SENSORTECHNICS GMBH ]
 浏览型号26PC10K0D2的Datasheet PDF文件第1页浏览型号26PC10K0D2的Datasheet PDF文件第3页浏览型号26PC10K0D2的Datasheet PDF文件第4页  
26PC Series
(mbar)
Temp. compensated and calibrated pressure sensors
PRESSURE SENSOR CHARACTERISTICS
V
s
= 10.0 ± 0.01 V, t
amb
= 20°C (unless otherwise noted)
Part number
26PC0050xxA
26PC0100xxA
26PC0250xxA
26PC01K0xxA
26PC02K0xxA
26PC05K0xxA
26PC10K0xxA
26PC16K0xxA
Operating pressure
0 - 50 mbar
0 - 100 mbar
0 - 250 mbar
0 - 1 bar
0 - 2 bar
0 - 5 bar
0 - 10 bar
0 - 16 bar
Min.
10.5 mV
20.0 mV
33.5 mV
93.0 mV
91.0 mV
69.0 mV
82.0 mV
133.0 mV
Full-scale span
2
Typ.
12.0 mV
24.0 mV
36.0 mV
97.0 mV
97.0 mV
72.0 mV
87.0 mV
140.0 mV
Max.
13.6 mV
27.0 mV
38.5 mV
100.0 mV
100.0 mV
76.0 mV
92.0 mV
147.0 mV
Sensitivity typ.
240 µV/mbar
240 µV/mbar
144 µV/mbar
97 µV/mbar
48.5 mV/bar
14.4 mV/bar
8.7 mV/bar
8.7 mV/bar
COMMON PERFORMANCE CHARACTERISTICS
V
s
= 10.0 ± 0.01 V, t
amb
= 25°C (unless otherwise noted)
Characteristics
Zero pressure offset
Temperature effects (0 - 50°C)
4
Offset
all 5 bar devices
all other devices
all 50 to 1000 mbar devices
all 2 bar devices
all other devices
all 50 and 100 mbar devices
all other devices
all 50 to 1000 mbar devices
all 2 bar devices
all other devices
Min.
-2.0
-1.5
Typ.
Max.
+2.0
+1.5
Unit
Span
Linearity (P2 > P1, BSL)
3
Repeatability and hysteresis
5
Long term stability
7
Input impedance
Output impedance
Response time
6
5.5
1.5
±0.5
±0.75
±1.0
±1.0
±0.75...±1.0
±0.25
±0.1
±0.1
±0.2
±0.5
7.5
2.5
±1.0
±1.5
±2.0
±2.0
±1.5
±0.5
±0.2
±0.7
mV
% sp a n
11.5
3.0
1.0
kΩ
ms
Specification notes:
1.
The maximum specified pressure which may be applied to the sensor without causing a permanent change in the output characteristics.
2.
Span is the algebraic difference between the output voltage at full-scale pressure and the output at zero pressure. Span is
ratiometric to the supply voltage.
3.
Linearity (BSL), the deviation of measured output at constant temperature (25°C) from "Best Straight Line" determined by three
points, offset pressure, full-scale pressure and half full-scale pressure.
V
- V
offset
[
V
�½ full scale
{
(full
full scale
pressure)
scale
x (�½ full scale pressure) + V
offset
}]
: 2 (V
full scale
) x 100 %
where: V = measured value for each device
4.
Error band of the offset voltage, span or bridge impedance in the specified temperature range, relative to the 25°C reading.
5.
Repeatability, the deviation in output readings for successive application of any given input pressure (all other conditions
remaining constant. Hysteresis, the error defined by the deviation in output signal obtained when a specific pressure point is
approached first with increasing pressure, then with decreasing pressure or vice versa (all other conditions remaining constant).
6.
Response time for 0 to full-scale pressure step change, readings taken at 10 % and 90 % of full-scale pressure.
7.
Long term stability of offset and span over a period over one year.
8.
The sensors might be used on both ports, for media compatible with the components, specified under "Service" (page 1).
9. Other sealing materials are available on request.
Minimum order quantities might be required.
10. Other pressure port styles, like barbed ones, luers, modular, M5, needle style or flow through connection, are
available on request.
Minimum order quantities might be required.
2/4
June 2005 / 596
www.sensortechnics.com