ELECTRICAL CHARACTERISTICS (CONT.)
HIGH LINEARITY VERSIONS
(VS+ = 5 V; Ta = 25° )
C
Parameter
Operating pressure range
MS5401-BM
Full-scale span (FS)
Sensitivity
Linearity
Operating pressure range
MS5405-BM
Full-scale span (FS)
Sensitivity
Linearity
Operating pressure range
MS5412-BM
Full-scale span (FS)
Sensitivity
Linearity
Operating temperature range
Zero pressure offset
Pressure hysteresis
All Ranges
Temperature hysteresis
Repeatability
Bridge resistance
Temperature coefficient of resistance
Temperature coefficient of span
Temperature coefficient of offset
Min
0
120
120
-
0
120
24
-
0
120
10
-40
-40
-
-
-
3.0
+2’400
-1’500
-80
Type
-
150
150
±0.05
-
150
30
±0.05
-
150
12.5
±0.05
-
0
-
0.3
-
3.4
2’900
-1’900
-
Max
1
180
180
±0.2
5
180
36
±0.15
12
180
15
±0.15
125
40
±0.20
0.8
±0.20
3.8
+3’300
-2’300
+80
Unit
bar
mV
mV/bar
% FS
bar
mV
mV/bar
% FS
bar
mV
mV/bar
% FS
°
C
mV
% FS
% FS
% FS
kΩ
ppm/°C
ppm/°C
µV/°C
Notes
8
1, 6
8
1, 6
8
1, 6
2, 6
3, 6
4, 6, 7
5, 6
5, 6
5, 6
NOTES
1) Deviation at one half full-scale pressure from the least squares best line fit over pressure range.
2) Maximum difference of output voltage after 1 pressure cycle at any pressure within the operating pressure
range.
3) Maximum difference in offset after one thermal cycle from -40° to +125°
C
C.
4) Same as 2) after 10 pressure cycles.
5) Slope of the end-point straight line from 25° to 60°
C
C.
6) Not 100% tested.
7) MS5412-BM: Max. 0.3% FS
8) This sensor family is optimised for the linearity; it is suitable for applications with higher pressure where the
linearity requirement is less critical.
DA54XX_011
000054xx280 ECN 927
May 9th, 2007
4