DC/DC CONVERTER 1W,
Reinforced Insulation, Medical Safety
MAU400 SERIES
Isolation, Safety Standards
I/O Isolation Voltage (rated)
I/O Isolation Test Voltage
I/O Isolation Resistance
Parameter
Flash tested for 1 Second
100KHz, 1V
500 VDC
60 Seconds
Conditions
4500
10
3000
Min.
Typ.
---
---
Max.
---
---
VACrms
G
V
PK
pF
Unit
I/O Isolation Capacitance
Safety Standards
Approvals
---
15
cUL/UL60950-1, CSA C22.2 No. 60950-1-03
UL60601-1,CSA C22.2 No.601-1
IEC/EN 60950-1, IEC/EN 60601-1
IEC60950-1 CB report, cUL/UL 60950-1 certificate
UL60601-1 UL certificate
---
20
---
General Specifications
Switching Frequency
MTBF (calculated)
Parameter
Conditions
MIL-HDBK-217F@25℃, Ground Benign
Min.
2,000,000
50
Typ.
80
---
Max.
100
---
Hours
KHz
Unit
Input Fuse
500mA Slow-Blow Type
5V Input Models
200mA Slow-Blow Type
12V Input Models
Environmental Specifications
Case Temperature
Operating Ambient Temperature Range (See Power Derating Curve)
Storage Temperature Range
Humidity (non condensing)
Cooling
Parameter
Conditions
Ambient
Min.
-25
---
---
Max.
+90
95
+85
Unit
℃
℃
% rel. H
℃
℃
-50
Free-Air convection
---
+125
Lead Temperature (1.5mm from case for 10Sec.)
260
Power Derating Curve
Notes
1
2
3
4
5
6
7
Specifications typical at Ta=+25℃, resistive load, nominal input voltage and rated output current unless otherwise noted.
Ripple & Noise measurement bandwidth is 0-20MHz.
These power converters require a minimum output loading to maintain specified regulation, operation under no-load conditions will not damage these
modules; however they may not meet all specifications listed.
All DC/DC converters should be externally fused at the front end for protection.
Other input and output voltage may be available, please contact factory.
That “natural convection” is about 20LFM but is not equal to still air (0 LFM).
Specifications are subject to change without notice.
E-mail:sales@minmax.com.tw
Tel:886-6-2923150
2013/01/24 REV:10
Page 2 of 4
Output Power (%)