Starter System Setup
shows how to connect the Multi-Cal-Power cable to power supplies and to the DMM. This
connection is for current loop output sensor modules. Refer to
for an illustration of how to set the
jumpers on the Multi-Cal-Test PCA for current loop configuration.
NOTE:
It is very important that the device power supply is floating. In other words, the negative
terminal on the device power supply is not connected to GND on the
±15-V
supply. Also, this
supply should not be referenced to earth ground.
Sense Input
HI
HI
LO
I
HP34401
DMM
LO
ImN
VmP
ImP
Linear
Power Supply
Floating Linear
Power Supply
VmN
Voltmeter
plugs are
not connected
+15V
GND
-15V
POS
24V
NEG
15vP
GND
15vN
J9
Multi-Cal-Master-PCA
Figure 15. Connect Multi-Cal-Power Cable to Power Supplies (Current Loop Output)
16
Multi-Cal-System Evaluation Module
Copyright
©
2010–2011, Texas Instruments Incorporated
SBOU087A
–
August 2010
–
Revised April 2011